Installation and Commissioning
Connecting a Cooling System
REGATRON
Performance. Precision. Quality.
TC.ACS - Device Manual (V04.60) - 26.4.2022
33 / 66
4.5.2.2 Requirements to the Cooling Liquid
In general the quality of normal purified, soft and largely chlorine-free drinking water is adequate. The quality of
the cooling medium will have a long-term effect on the system performance.
CAUTION
Possible damage to the device!
- Due to electrolytic and/or chemical corrosion
- Due to the deposition of fine sludge
- Due to coating with foreign substances and as a result degradation of the heat transfer
- Due to the deposition of foreign bodies on the heat exchanger surfaces
Avoidance:
- Never use deionised water as a cooling liquid.
- Use fine filters to keep back fine sludge particles.
- Meet the following requirements.
Requirements to the Cooling Liquid
Parameter
Unit
Limit value
ph-value
-
6 - 8
Total hardness
[°dH]
2
< 15
Carbonate hardness
[°dH]
2
< 6
Non-carbonate hardness
[mmol/l]
3
< 1.60
Carbon dioxide, free
[mg/l]
< 3
Organic substances
[mg/l]
< 10
Algae and fungi
Not allowed
Sand and sludge
Not allowed
Sulphate SO
3
[mg/l]
< 50
Chlorine CL
[mg/l]
< 30
Undissolved iron Fe
[mg/l]
< 1
Phosphate P
2
O
5
Not allowed
Total salt content
[mg/l]
< 3000
Manganese Mn
[mg/l]
< 0.1