REED ST-615 Скачать руководство пользователя страница 3

reedinstruments

www

com

Pantone 485 & 123
485 = (0/100/100/10)
123 = (0/18/100/0)

Specifications

Display:..

Dual.LCD.with.Backlight.and.status.Indicators

Sensor.Type:...
..

1 - Humidity: 

.

Precision.capacitance.sensor

Function

Range &  

Resolution

Accuracy

Humidity

5.0.to.98.0%.RH

±.3.5%.RH.

Temperature.(K-type).

(probe.accuracy.not.

included)

-20.0.to.199.9°C

±(3.0%.r.4°C)

-20.to.1000°C

-4.0.to.199.9°F

±(3.0%.r.7°F)

-4.to.1832°F

.

 2 - Temperature:

..

Thermistor.(probe).and.type.“K”.thermocouple.

Response.Time:.

Probe.Temperature.and..

.

Relative.Humidity:.3.minutes

Accuracy note: 

Accuracy is specified for the following ambient 

.

.

temperature.range:.64.to.82°F.(18.to.28°C)

Sampling.Rate:.

2.5.samples.per.second

Operating.Conditions:. 32.to.122°F.(0.to.50°C);..
.

<.80%.RH.non-condensing

Storage.Conditions:.. 14.to.140°F.(-10.to.60°C);..
.

<80%.RH.non-condensing

Power.Supply:..

9V.Battery.with.auto.power.off.(after.10.minutes)

Dimensions:.

5.9.x.2.8.x.1.4”.(150.x.72.x.35mm)

Weight:..

210g

Содержание ST-615

Страница 1: ...reedinstruments www com Pantone 485 123 485 0 100 100 10 123 0 18 100 0 Model ST 615 Digital Thermo Hygrometer with remote or integral probe Instruction Manual...

Страница 2: ...us numerous status indicators Careful use of this meter will provide years of reliable service Features Semiconductor sensor for temperature measurements of 0 C to 40 C and 32 F 104 F plus type K ther...

Страница 3: ...eading 7 F 4 to 1832 F 2 Temperature Thermistor probe and type K thermocouple Response Time Probe Temperature and Relative Humidity 3 minutes Accuracy note Accuracy is specified for the following ambi...

Страница 4: ...surement reading display 5 Power ON OFF button 6 Thermometer measurement buttons 6 7 Relative Humidity function buttons 2 8 Backlight function buttons Display Descriptions Probe temperature F C measur...

Страница 5: ...mperature in F set the function switch to the F range If you wish to measure temperature in C set the function switch to the C range 2 Insert the Temperature Probe into the Temperature Socket making s...

Страница 6: ...or 1 F Data Hold Buttons 1 Displays can be held frozen at any time by pressing the HOLD button For Temperature use the HOLD button on the left For Relative Humid ity use the HOLD button on the under...

Страница 7: ...ght of the LCD when the 9V battery needs to be replaced Replace the 9V battery as follows 1 Turn the meter off and disconnect the air temperature probe 2 Remove the large flat head screw on the back o...

Страница 8: ..._____________ ________________________________________________ ________________________________________________ ________________________________________________ _______________________________________...