Features

•  Compact single input thermocouple thermometer designed for field use 
•  Accuracy to 0.5% 
•  Display maximum reading plus Data Hold 
•  Large LCD display with backlight 
•  User selectable °C or °F or Kelvin
•  User selectable 0.1°C or 1°C (0.1°F or 1°F) resolution 
•  Accepts K-type thermocouples (see page 4 for selection)
•  Includes 9V battery and holster 

Specifications

Electrical

Temperature Scale: 

Celsius (C°), Fahrenheit (°F) or Kelvin (K) 

Measurement Range: 

-50°C to 1300°C, -58°F to 2000°F, 223K to 2000K

Resolution: 

1°C or 1°F, 0.1°C or 0.1°F,1 K

Accuracy.

Accuracy is specified for operating temperatures 

.

 

over the range of 18°C to 28°C (64°F to 82°F), 

.

 

for 1 year, not including thermocouple error

 

±2°C 

-50°C to 0°C

 

±4°F 

-58°F to 32°F

 

±5K 

223K to 273K

 

±(0.5% rdg+1°C) 

0°C to 1000°C

 

±(0.8% rdg+1°C) 

1000°C to 1300°C

 

±(0.5% rdg+2°F) 

32°F to 2000°F

 

±(1.0% rdg+2K) 

273K to 2000K

Temperature Coefficient:  0.1 times the applicable accuracy specification 

.

 

per °C from 0°C to 18°C and 28°C to 50°C 

.

 

(32°F to 64°F and 82°F to 122°F)

Input protection: 

60V DC or 24V RMS AC maximum input 

.

 

voltage on any combination of input pins

Reading Rate:  

2.5 times per second

Input Connector:  

Accepts standard miniature thermocouple 

.

 

connectors (flat blades spaced 7.9mm, 

.

 

center to center)

continued ...

Содержание ST-610B

Страница 1: ...reedinstruments www com Model ST 610B Digital Thermocouple Thermometer Instruction Manual...

Страница 2: ...discon nect the thermocouple connectors from the thermometer before remov ing the cover To avoid damage or burns do not make temperature measurements in microwave ovens Cautions Repeated sharp flexin...

Страница 3: ...or 1 F 0 1 C or 0 1 F 1 K Accuracy Accuracy is specified for operating temperatures over the range of 18 C to 28 C 64 F to 82 F for 1 year not including thermocouple error 2 C 50 C to 0 C 4 F 58 F to...

Страница 4: ...139 Surface Probe Spring loaded probe is ideal for all high temperature applications Model LS 104 Right Angle Surface Probe Unique right angle design allows temperature of hard to reach areas to be t...

Страница 5: ...ent range selected If measuring above 199 9 change the resolution to 1 Be certain to seat the thermocouple connector properly and that the leads are not broken Hold Mode Pressing the HOLD button to en...

Страница 6: ...emperature Battery Replacement WARNING To avoid possible electric shock disconnect the thermocouple connectors from the thermometer before removing the cover The low battery symbol appears on the lowe...

Страница 7: ..._____________ ________________________________________________ ________________________________________________ ________________________________________________ _______________________________________...

Отзывы: