Calibration
1. Open the cover plate and place 2-3 drops of distilled water onto the prism
surface, close the cover plate, press lightly and ensure the water spreads
evenly across the whole surface area without air bubbles or dry sections.
(wait 30 seconds until the water adjusts to ambient temperature)
2. Hold towards light source and look into the eyepiece, if the boundary line
between the upper blue field and the lower white field is not located exactly
on the 0°C waterline, adjust the correcting screw until both lines meet.
3. Wipe the water from the prism with clean dry cloth and replace the rubber
correcting screw cover.
NOTE
: under normal operating conditions the calibration only needs to be checked
periodically.
Maintenance
1.
If the ambient temperature is not within the automatic temperature
compensation range (10 - 30°C) the instrument must be calibrated before
testing commences.
2.
The prism must be cleaned thoroughly with a soft, clean damp cloth after
each use - failure to do so will result in inaccurate readings and possible
long term
damage.
3.
The instrument must be kept in a dry clean environment to prevent
condensation forming on the optical components.
4.
Use a glass rod to sample battery fluid as acid will damage the dropper and
also harm your skin–separate readings must be taken from each cell
5.
If used in accordance with the instructions and all warning are followed -
the optical performance will remain constant and the instrument will give
many years of reliable service.