Troubleshooting
Error messages are displayed by a number code on the LCD.
When the Hardware Error icon appears on the LCD, turn the meter on and
off several times. If the symbol still appears contact REED Instruments at
[email protected] for repair.
Be sure to take care with the meter’s mirror. Avoid dirt, scratches, and
solvents. These may affect the accuracy and precision of the meter.
Code Cause
Remedy
101
Battery too low
Change batteries
104
Calculation Error
Repeat procedure
152
Temperature too high
Cool down the meter
153
Temperature too low
Warm up the meter
154
Out of range
Put target between 0.05 & 45m
155
Signal too weak
Ensure surface is flat and
large enough
156
Signal too strong
157
Measure error or background
brightness
Darken or change target
160
Too much shaking
Stabilize the meter
www.reedinstruments.com
1-877-849-2127 | [email protected]