Operating Instructions
1. Power on the instrument by pressing the Power Button.
2. Position the Light Sensor directly under the light source. Mean-
while the light reading value will be displayed on the LCD display.
3. Press the Lux/Ft-cd Button to select measuring unit Lux or Ft-cd.
Zero Offset Adjustment of Light Function
•
For best results you should zero the light sensor prior to use in a dark
environment. To achieve this, place the light sensor end of the meter
under a desktop or flat surface to block any light. Press the Unit/Zero
Button setting the meter indication to zero.
•
The zero point can drift due to environment temperatures, battery
power, as well as various other reasons. It is recommended that the
zero be checked frequently using the above procedure.
Hold Function
Pressing the Hold Button) will freeze the current value with a “HOLD”
symbol on the display. Press again to release the hold function.
Data Record Function
The Data Record function records & displays the maximum and minimum
reading values.
1. Press the Max/Min Button once and the “Max” symbol along with
the maximum value will appear on the display.
2. Press the Max/Min Button again, the “Min” symbol along with the
minimum value will appear on the display.
3. To exit the memory record function, press the Max/Min Button
continuously for at least 2 seconds. The display will revert to the
current reading.
4. Clear the Max/Min value recorded by pressing the Hold Button.
Previous recorded Max/Min value will be given up and then revert
to the REC. function keep on recording.