Maintenance

Cleaning the Test Piece

After taking a measurement, clean the test pieces to prevent them from 
rusting. If the pieces are not being used for a long period of time coat 
them with oil to prevent rust.

Protecting the Ultrasonic Sensor

Be sure to clean the ultrasonic sensor and cable after each use. Grease, oil 
and dust will cause the cable to dry out and shorten life expectancy. The 
temperature of the surface being measured should not exceed 140°F (60°C).

Replacing the Ultrasonic Sensor

The degradation and wear of the probe's interlayer plate will influence 
measurements. Replace the probe when the following occurs:
1. 

If it always displays the same value when measuring different thicknesses

2.  When a measurement displays no value

Battery Replacement

1.  When the low battery symbol appears on the display, replace  

the batteries.

2.  Remove the battery cover on the back and insert 4 x "AAA" batteries.

Applications

•  Monitoring and verifying pipes and pressure vessels
•  Industrial manufacturing

www.

.com

[email protected]

1.800.561.8187

Содержание TM-8811

Страница 1: ...www REEDINSTRUMENTS com Ultrasonic Thickness Gauge Instruction Manual TM 8811 www com information itm com 1 800 561 8187...

Страница 2: ...Auto Power OFF 8 Setting the Sound Velocity for a Known Material 8 10 Determining the Sound Velocity for a Material with a Known Thickness 10 Measurement Methods 11 Maintenance 12 Cleaning the Test Pi...

Страница 3: ...ED distributor or authorized Service Center Please note an additional fee for this service will apply Safety Never attempt to repair or modify your instrument Dismantling your product other than for t...

Страница 4: ...d nodular cast iron Sampling Time Less than 1 second Display 4 Digit LCD Probe Length 3 36 Low Battery Indicator Yes Power Supply 4 x AA Batteries Product Certifications CE Operating Temperature 32 to...

Страница 5: ...3 Power Button 4 Up Arrow 5 Calibration Test Block 6 Probe Connectors 7 Coupling Indicator 8 LCD Display 9 Coupling LED 10 Calibration Button 11 Unit of Measure Button 12 Down Arrow 13 Velocity Butto...

Страница 6: ...ss CAL again to save 6 Calibration is now complete and the ultrasonic sensor can now be removed from the calibration test block Note Results from the calibration procedure will automatically save in t...

Страница 7: ...5 Brass 6 Cd06 Zinc 7 Cd07 Quartz Glass 8 Cd08 Polyethylene 9 Cd09 PVC 10 Cd10 Gray Cast Iron 11 Cd11 Nodular Cast Iron 12 XXXX User Defined Sound Velocity 4 Press the or arrows to select the desired...

Страница 8: ...r the measured values appear to be erratic verify that there is an adequate amount of coupling gel in between the ultrasonic sensor and the material under test It is also important that the ultrasonic...

Страница 9: ...7 The sound velocity can be adjusted by pressing the or arrows to match the sound velocity for the material under test as shown in the table below Material Sound Velocity m s inch s Aluminum 6320 to...

Страница 10: ...rming the sound velocity has been set If you set a sound velocity without confirming the selection the meter will automatically resume normal operation after a few seconds Determining the Sound Veloci...

Страница 11: ...s near a single spot rotating the ultrasonic sensor from 0 to 90 respectively with respect to the split face Fig 1 Take the smaller of the two indicated values as the thickness of the material 3 Multi...

Страница 12: ...f the surface being measured should not exceed 140 F 60 C Replacing the Ultrasonic Sensor The degradation and wear of the probe s interlayer plate will influence measurements Replace the probe when th...

Страница 13: ...on the instrument Use on external parts only Product Warranty REED Instruments guarantees this instrument to be free of defects in material or workmanship for a period of one 1 year from date of shipm...

Страница 14: ...ease contact your authorized REED distributor or REED Instruments Customer Service by phone at 1 877 849 2127 or by email at info reedinstruments com Please visit www REEDINSTRUMENTS com for the most...

Страница 15: ...www REEDINSTRUMENTS com TEST MEASURE WITH CONFIDENCE CHECK OUT OUR LATEST PRODUCTS www com information itm com 1 800 561 8187...

Страница 16: ...www REEDINSTRUMENTS com www com information itm com 1 800 561 8187...

Отзывы: