background image

Sound  

Level Meter/ 

Data Logger

Instruction 

Manual

R8080

 

REED Instruments

1-877-849-2127 | [email protected] | www.reedinstruments.com

Содержание R8080

Страница 1: ...Sound Level Meter Data Logger Instruction Manual R8080 REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com...

Страница 2: ...F 8 Backlight 8 Setting the Date and Time 8 Setting the Datalogging Recording Interval 9 Datalogging 9 Clearing Stored Data 9 Calibration Procedure 9 10 Battery Replacement 10 AC DC Signal Output Jack...

Страница 3: ...ce will apply Safety Never attempt to repair or modify your instrument Dismantling your product other than for the purpose of replacing batteries may cause damage that will not be covered under the ma...

Страница 4: ...condenser Digital Analog Bargraph Yes 30 segment Display 4 digit LCD Backlit Display Yes Min Yes Max Yes Alarm Indicators Under and Over visual on screen Datalogging Capabilities Yes Real Time Clock a...

Страница 5: ...USB Cable PC Software Hard Carrying Case Batteries Instrument Description 1 Windscreen 2 Display 3 Power Backlight Button 4 MIN MAX Clock Button 5 Record Button Up 6 Frequency Weighting Setup Button...

Страница 6: ...M drive from your computer Double click on the setup file and follow the on screen prompts to install the software If your computer does not have a CD drive please visit www reedinstruments com softwa...

Страница 7: ...oximately in the middle of the bar graph If the over range indicator appears during your measurement the upper limit of the selected dB range has been exceeded Adjust the selected dB range until the o...

Страница 8: ...ed to a PC Backlight Press the Backlight Button to turn the LCD Backlight on and off The backlight will automatically turn off after 30 seconds Setting the Date and Time In order to set the time and d...

Страница 9: ...T SLOW INTV button to save the desired sampling rate and resume normal operation Note Press the Power button to exit Setup mode at any time Datalogging 1 Turn the meter on and select your desired samp...

Страница 10: ...cated on screen by CAL 94dB 4 Insert the microphone into the calibrator 5 Press the Up and Down arrow buttons to increase or decrease the dB level to match the Sound Calibrator output 6 Once the desir...

Страница 11: ...Connector Diagram AC Output Voltage Range 1 Vrms at FS full scale FS refers to the upper limit of the selected dB level range For example if 30 90dB is the selected level range 1Vrms 90dB DC Output V...

Страница 12: ...Instruments guarantees this instrument to be free of defects in material or workmanship for a period of one 1 year from date of shipment During the warranty period REED Instruments will repair or repl...

Страница 13: ...ED distributor or REED Instruments Customer Service by phone at 1 877 849 2127 or by email at info reedinstruments com Please visit www REEDINSTRUMENTS com for the most up to date manuals datasheets p...

Страница 14: ...Test Measure with Confidence Check out our latest Products REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com...