
6
When single mode is selected, the measurement indicator
will only
appear at the top right of the LCD when a measurement will be taken by
the user (see
Taking Single Measurements
section for details).
Measuring Geometry
Gloss is classed as either low, semi or high gloss. In order to determine
the best measuring geometry for your test start at a 60° angle as it is
considered the universal measurement angle.
The angle is selected based on the anticipated gloss range, as shown in
the table below.
Gloss Range
60° Value
Notes
High Gloss
>70 GU
If measurement exceeds 70 GU,
change test setup to 20°.
Medium Gloss
10 to 70 GU
Low Gloss
<10 GU
If measurement is less than 10 GU,
change test setup to 85°
Taking Single Measurements
1. Place the measurement aperture against the material.
2. Press the
ANGLE
button to select the right measuring geometry.
3.
Take measurements by pressing the
MEAS
button.
4. The reading that displays within 2 seconds is the gloss value at the
selected angle position.
5. Each result is automatically logged into memory.
Note:
Find the center of measurement by the intersection of the arrows
marked on the front and side of the instrument.
Taking Continuous Measurements
1. Place the measurement aperture against the material.
2. Press the
ANGLE
button to select the right measuring geometry.
3. The reading that displays within 2 seconds is the gloss value at the
selected angle position.
continued...
www.
.com
1.800.561.8187