11
• Pay particular attention not to connect the current signal to the input
terminal. Incorrect connection may cause damage to the meter and the
instrument under test.
Measuring DC Voltage
1. Insert one end of the test lead into the INPUT terminal of the meter
and connect the other end to the output terminal of the instrument
under test, as shown in the diagram to the right.
2. Press
INPUT/OUTPUT
button to
select the Input function. 'INPUT',
'mV' will be displayed.
3. Starts the measurement process by
displaying '000.00' which indicates
waiting and then displays measurement
result. (Refresh rate is twice a second,
and if the measured value exceeds the
measurement range, the LCD
displays 'OL'.)
Measuring Thermocouples
1. Insert one end of the test lead into the INPUT terminal of the meter
and connect the other end to the output terminal of the instrument
under test, as shown in the diagram above.
2. Press
INPUT/OUTPUT
button to select the Input function.
3. Press the
FUNC
button to select thermocouple function.
'R' and '°C' will appear on the display.
4. Press the
RANG
button to select the type of thermocouple.
5. Starts the measurement process by displaying '000.00' which
indicates waiting and then displays measurement result.
(Refresh rate is twice a second, and if the measured value exceeds
the measurement range, the LCD displays 'OL'.)
www.
.com
1.800.561.8187