background image

www.REEDINSTRUMENTS.com

Infrared Temperature 

Calibrator

Instruction 

Manual

BX-500

1.888.610.7664 

[email protected]

REED-Direct.com

Содержание BX-500

Страница 1: ...www REEDINSTRUMENTS com Infrared Temperature Calibrator Instruction Manual BX 500 1 888 610 7664 info REED Direct com REED Direct com ...

Страница 2: ...uctions 9 12 Initial Start Up 9 Turning Off the Calibrator 10 Selecting Temperature Unit of Measure 10 Setting the Temperature Set Point 10 Setting Overload Alarms 1 2 11 Enabling Disabling the Heating Power 11 Keypad Lock Unlock 12 Replacing the Fuses 12 Product Care 13 Accessories and Replacement Parts 13 Applications 13 Product Warranty 13 Product Disposal and Recycling 14 Product Support 14 1 ...

Страница 3: ...orized Service Center Please note an additional fee for this service will apply Safety This instrument can generate extreme temperatures Precautions must be taken to prevent personal injury or damage to objects To avoid personal injury read and follow these guidelines Never attempt to repair or modify your instrument Dismantling your instrument other than for the purpose of replacing fuses may cau...

Страница 4: ...ls Use of this instrument at high temperatures for extended periods of time requires caution Components and heater lifetime can be shortened by continuous high temperature operation Do not use fluids to clean the target surface Features For calibrating non contact temperature instruments infrared thermometers pyrometers and thermal imagers Wide aperture compensates for optical infrared thermometer...

Страница 5: ... 0 4 C Resolution 0 1 F 0 1 C Target Emissivity 0 95 Fixed Heating Time Approximately 30 mins to maximum Cooling Time Approximately 30 mins to 122 F 100 C Aperture Diameter 2 28 58mm Display Dual LED Power Supply 110VAC 2 5A Product Certifications CE Operating Temperature 41 to 95 F 5 to 35 C Storage Temperature 4 to 140 F 20 to 60 C Operating Humidity 15 to 80 Dimensions 7 1 x 4 5 x 8 8 180 x 114...

Страница 6: ...lowing conditions Temperature range 41 to 95 F 5 to 35 C Ambient relative humidity 15 to 80 Pressure 75 to 106kPa Mains voltage within 10 of nominal Temperature difference between the target surface and the RTD PT100 reference well located directly behind the target surface are calculated as shown in the graph below 1 888 610 7664 info REED Direct com REED Direct com ...

Страница 7: ... Indicators 4 Set Button 5 Return Button 6 Down Arrow 7 Up Arrow 8 Infrared Target Surface 9 RTD Reference Well For Calibration Purposes Only 10 Fan 11 Heating System Fuse 12 Control System Fuse 13 Power Socket 14 Power Switch 1 2 3 4 8 9 5 6 7 BX 500 13 14 11 12 10 continued 1 888 610 7664 info REED Direct com REED Direct com ...

Страница 8: ...rmance and could damage the calibrator AT Auto Tuning Parameter Adjustment For Authorized Service Centers Only OUT Heating Output Power Indicator ALM1 Overload Alarm indicates that the IR Calibrator temperature exceeds the set temperature To set the alarm refer to Setting Overload Alarms 1 2 for details Note When triggered the heating output power is turned off ALM2 Overload Alarm indicates that t...

Страница 9: ...meter Operating Instructions Initial Start up 1 Place the calibrator on a flat surface with at least 8 of free space around the instrument 2 Connect the included power cord and plug into a power outlet 3 Turn the calibrator on by toggling the power switch to ON 4 The fan should begin blowing air through the instrument 5 After a brief 3 second self test the calibrator will begin normal operation by...

Страница 10: ...ates At oFF confirming that the parameters are now unlocked 2 Press the SET button for three seconds and release to enter the setup mode as indicated by and 3 Press button once and the display will indicate and which confirms the current temperature scale 4 Use the and arrows to select between C and F 5 Press the SET button to confirm selection 6 Press the SET button again to exit set up mode and ...

Страница 11: ...igh overload alarm 1 4 Press the button once to switch to high overload alarm 2 as indicated by 5 Use the and arrows to change the temperature value which will flash until confirmed 6 Press the SET button to confirm the new high overload alarm 2 7 Press the SET button again to exit set up mode and resume normal operation Enabling Disabling the Heating Power 1 Press the button once to display the h...

Страница 12: ...gain to exit set up mode and resume normal operation Replacing the Fuses The calibrator must be switched off before attempting to replace fuses To replace fuses 1 Open the faulty fuse compartment by rotating it counter clockwise 2 Gently slide out the fuse 3 Replace with the appropriate fuse see Specifications below and tighten the compartment back into place Fuse Specifications Fuse 1 3A 250V Fus...

Страница 13: ...d Thermometers and Thermal Imagers Product Warranty REED Instruments guarantees this instrument to be free of defects in material or workmanship for a period of one 1 year from date of shipment During the warranty period REED Instruments will repair or replace at no charge products or parts of a product that proves to be defective because of improper material or workmanship under normal use and ma...

Страница 14: ...e contact your authorized REED distributor or REED Instruments Customer Service by phone at 1 877 849 2127 or by email at info reedinstruments com Please visit www REEDINSTRUMENTS com for the most up to date manuals datasheets product guides and software Product specifications subject to change without notice All rights reserved Any unauthorized copying or reproduction of this manual is strictly p...

Страница 15: ...www REEDINSTRUMENTS com TEST MEASURE WITH CONFIDENCE CHECK OUT OUR LATEST PRODUCTS 1 888 610 7664 info REED Direct com REED Direct com ...

Страница 16: ...www REEDINSTRUMENTS com 1 888 610 7664 info REED Direct com REED Direct com ...