Reed Instruments 800837005953 Скачать руководство пользователя страница 12

12

2.  Press the 

AL

 and 

SET

 buttons to scroll through the stored values. 

3. 

To clear the stored values, press and hold the MEM button first and then 

press the 

 button.

Note:

 If no values are stored or if the memory is cleared the display will 

appear as shown below.

4.  Press the MEM button to exit memory recall mode and resume 

normal operation.

Setting the Critical Alarms 

1.  Press the 

AL

 button to turn the critical alarm mode ON or OFF.

2.  Hold the 

SET

 button for approximately 3 seconds to enter the critical 

alarm set up mode.

3.  Press the 

SET

 button to toggle between digits.

4.  Press the 

AL

 and 

MEM

 buttons to adjust the flashing digit.

Note:

 The maximum set critical alarm value that can be set is 199Ω/499mA.

5.  Press the 

SET

 button to confirm your selection and proceed to the 

next digit.

6.  When complete, hold the 

SET

 button for approximately 3 seconds to 

exit the critical alarm set up mode and resume normal operation.

Note:

 When the critical alarm is enabled, the instrument will begin to beep 

if the measured vale exceeds the selected alarm.

7.  Follow steps 2 to 6 to set the critical alarm in mA.

 

REED Instruments

1-877-849-2127 | [email protected] | www.reedinstruments.com

Содержание 800837005953

Страница 1: ...Clamp On Ground Resistance Tester Instruction Manual R5710 REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com...

Страница 2: ...er ON OFF 8 Auto Power OFF 9 Resistance Ground Leakage Measurement Mode 9 Resistance Measurement Mode 9 10 Data Hold Memory Storage 11 Memory Recall 11 12 Setting the Critical Alarms 12 Battery Replac...

Страница 3: ...r the manufacturer s warranty Servicing should only be provided by an authorized service center Be aware of your surroundings environment and measurement scope If the instrument applies to ground resi...

Страница 4: ...res up to 300 readings User adjustable alarms Data hold function Durable double molded conductor clamp 1 26 32mm jaw size for large ground rods Low battery indicator and auto shut off Cat III 300V saf...

Страница 5: ...00 to 880 900 to 1200 Accuracy 0 010 to 0 099 1 0 01 0 10 to 0 99 1 0 01 1 0 to 49 9 1 0 1 50 0 to 99 5 1 5 0 5 100 to 199 2 1 200 to 395 5 5 400 to 590 10 10 600 to 880 20 20 900 to 1200 25 30 Resolu...

Страница 6: ...it LCD Yes User Selectable Alarms Yes Display Hold Yes Internal Memory Yes up to 300 data points Auto Shut off Yes after 5 minutes Power Supply 4 x AA Batteries Over Range Indicator Yes Low Battery In...

Страница 7: ...old Indicator 3 Battery Status Indicator 4 Resistance Measurement Indicator 5 Resistance Measurement Value 6 Current Measurement Indicator 7 Current Measurement Value 8 Number of Stored Values 9 Inter...

Страница 8: ...e for details Note The instrument automatically enters the resistance earth leakage measurement mode at start up regardless of the last saved setting Note DO NOT clamp on to any conductor or open the...

Страница 9: ...earth leakage simultaneously 2 Pull the trigger to open the clamp and clamp onto the circuit under test 3 The LCD will display the resistance and ground leakage current measurement values 4 If the in...

Страница 10: ...ill display the resistance value 5 If the instrument beeps and the LCD display flashes the resistance measurement is exceeding the critical set alarm Refer to Setting the Critical Alarms for details c...

Страница 11: ...ss the button again to resume normal operation Note When the Data Hold feature is active the instrument will not turn OFF 4 Repeat steps 1 to 3 to hold store new values during testing Note When the me...

Страница 12: ...nter the critical alarm set up mode 3 Press the SET button to toggle between digits 4 Press the AL and MEM buttons to adjust the flashing digit Note The maximum set critical alarm value that can be se...

Страница 13: ...defects in material or workmanship for a period of one 1 year from date of shipment During the warranty period REED Instruments will repair or replace at no charge products or parts of a product that...

Страница 14: ...ED distributor or REED Instruments Customer Service by phone at 1 877 849 2127 or by email at info reedinstruments com Please visit www REEDINSTRUMENTS com for the most up to date manuals datasheets p...

Страница 15: ...www REEDInstruments com TEST MEASURE WITH CONFIDENCE CHECK OUT OUR LATEST PRODUCTS REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com...

Страница 16: ...www REEDInstruments com REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com...

Отзывы: