Redsense Medical Alarm Unit Скачать руководство пользователя страница 4

4

IFU | 

Alarm unit,  Fiber optic cable & Sensor patch | 

ENGLISH | 

© 

Indications For Use

The Redsense device is intended to monitor for
potential blood loss from the hemodialysis access
site in hemodialysis patients undergoing hemo-
dialysis treatment. The device includes a blood
sensor incorporated into a sensor patch. The sensor
monitors potential blood leakage from the venous
needle blood access via a light signal and will alarm
if blood leakage is detected by the device’s sensor.

Mounting Redsense

The Redsense sensor should be placed on the patient
after the needles have been inserted and before
dialysis begins. Redsense must be turned on the entire
time that dialysis is running.

1.

  Attach the alarm unit to the dialysis machine’s IV pole / drip stand 

(see figure 1)

. If this isn’t possible please contact

  Redsense Medical for consultation. Connect the AC adapter to a wall outlet and to the black-colored
  connector on the alarm unit. 

(Figure 2 shows the connector.)

2.

   The alarm unit indicates as follows whether the built-in battery is charged:

  •   A flashing battery indicator light means that the battery is being charged. The unit cannot be started in this   
    case. Wait until the charging is complete.
  •  A solid battery indicator light indicates that charging is complete and monitoring can begin.

3.

  Connect the optical extension fiber to the alarm unit. Turn the coupling roughly a 1/3 turn to lock it. Check

  that the coupling is locked 

(see figure 3).

4. 

Check the date on the packaging for the sensor patch and be sure that the packaging has a Redsense label on  

  it. Open and check that the sensor patch does not have any mechanical damage.

5.

  In order to start Redsense, push the on/off button and check that all the indicator lights are on and the 

  speaker is functional. A short beep should be heard. If it is the first set-up or it is a new environment for the    
  Redsense, please check that the patient’s supervising staff or observing person can clearly hear the alarm
  sound at the correct distance. The alarm signal should be a continuous signal with three different levels/tones
  and can be initiated by starting Redsense and then provoking  an alarm. The observing person must be sure
  that the alarm signal can be heard from other rooms or while there is noise. The observing person must be
  situated at places where the alarm signal can be heard during the entire dialysis treatment. If this isn’t possible
  please consult Redsense Medical.

6.

  To stop testing the indicator lights and the alarm signal, push the on/off button again. The green indicator light  

  starts to flash together with a intermittent signal.

7.

  Connect the fiber from the sensor patch to the optical extension fiber.  Open the connector on the end of the

  optical extension fiber by holding in the locking button, inserting the sensor patch fiber from the sensor patch  
  until it stops/resists, then releasing the lock button 

(see figure 4).

 When the green indicator light is solid, and the

  intermittent signal is turned off the monitoring function is working correctly.

8.

    Make sure the area around the needle is dry.

9.

    Apply the sensor patch to the skin over the needle with the absorbent part of the sticker centered directly

    over the insertion point. Be sure that the self-adhesive part of the patch does not cover the wings 

(see figure 5).

10.  

Secure the needles in place as usual using tape or similar material. Secure the needles in place according to

    the existing policy and procedure. Dressing material and or tape may be applied directly over the surface of    
    the sensor patch according to the facility policy and procedure.

Figure 6

 shows an example of how the sensor patch and the optical extension fiber can be attached to the

patient’s arm. Check that the sensor patch is clean and dry.

Содержание Alarm Unit

Страница 1: ...Instructions for use Alarm unit Fiber optic cable Sensor patch ENGLISH ...

Страница 2: ...edical com These instructions are released with the following date version 2017 01 18 Instructions RM 1RM240 Instructions Warnings and precautions 3 Mounting Redsense 4 During dialysis 6 After completing dialysis 6 Maintenance 6 Cleaning 6 Charging the battery 6 Replacing the battery 7 Symbols and terms 7 Technical Information 9 EMC Information 9 ...

Страница 3: ...radio communication equipment may affect Redsense Redsense is not intended to be used in a highly oxygenated environment Modifying the equipment is not permitted Modification of the equipment will void all guarantees and product liability on the part of the manufacturer Servicing of the equipment may only be performed by Redsense Medical or by persons authorized by them External equipment intended...

Страница 4: ... should be heard If it is the first set up or it is a new environment for the Redsense please check that the patient s supervising staff or observing person can clearly hear the alarm sound at the correct distance The alarm signal should be a continuous signal with three different levels tones and can be initiated by starting Redsense and then provoking an alarm The observing person must be sure t...

Страница 5: ...5 IFU Alarm unit Fiber optic cable Sensor patch ENGLISH Warnings and precautions figure 1 figure 2 1 2 figure 3 figure 4 1 2 YES YES NO figure 5 figure 6 ...

Страница 6: ...is not connected or has been broken The optical extension fiber is not connected correctly Turn off the alarm by keeping the on off button pressed fix the cause of the alarm and then restart the alarm unit The yellow warning indicator together with the battery indicator followed by an intermittent alarm signal means that the alarm unit is no longer receiving power from the AC adapter which means t...

Страница 7: ...pment and changing the battery should only be done using Redsense Medical or by personnel authorized by them Symbols and terms The following symbols and terms are used in these instructions and on the product Connection for optic fiber On off button Green indicator with operating symbol Yellow indicator with battery symbol Yellow indicator with warning symbol Red indicator with alarm symbol ith Co...

Страница 8: ... month YYYY MM On 0ff button Do not use if package is damaged Batchnumber includes manufacturing year Alarm indicator Item number On off indicator Serial number includes manufacturing year Warning indicator Consult the instruction for use Applied part of type BF Body Floating the surface between the equipment and the patient does not conduct electricity The alarm unit covered by the WEEE directive...

Страница 9: ...recautions Alarm unit 128 x100 x 35 mm 150 g 15 to 40 C 20 to 60 C 10 to 40 C 75 75 75 5 years See EMC Information Extension fiber Ø10mm x diffent lenghts ca 10 30 g 15 to 40 C 20 to 60 C 10 to 40 C 75 75 75 5 years Sensor patch 140 x 60 x 1 mm 150 g 15 to 40 C 20 to 60 C 10 to 40 C 75 75 75 See expiry date RF emissions CISPR 11 RF emissions CISPR 11 Harmonic emission IEC 61000 3 2 Voltage fluctua...

Страница 10: ...5 cycles 5 UT 95 dipinUT for 5 sec 3A m 3Vrms 3V m Electromagnetic environment Electromagnetic environment guidance Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or ho...

Страница 11: ... distances between portable an mobile RF communications equipment and Redsense Redsense is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of Redsense can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and Redsense as recom m...

Страница 12: ...ellow indicator with warning symbol Red indicator with alarm symbol on ator with ymbol Connector for AC adapter Service connector Redsense use only ellow indicator with attery symbol Yellow indicator with warning symbol or with ol Redsense Medical AB Gyllenhammars väg 26 302 92 HALMSTAD SWEDEN www redsensemedical com ...

Отзывы: