background image

 

RL-101U User Manual 

 

 

 

Powering ON/OFF and selecting Gas:   To power on sensor head click the 

Test

 

button until you see it show “Adr” on the display. Once you see “Adr” 

you will click the 

menu

 

button 3 times and it should show “OFF” or a type 

of gas(H2S/LEL/CO/SO2). Once you see one of those options click the 

UP

 

button until you see the correct choice for what you will be using. You will 
do these same steps to turn it off or select a different type of gas if needed.  

 

Revision Menu walk through by hitting the menu button after each mode.  

1.

 

RE/A24

” –

 This is the Revision mode of the menu, it just tells you which 

revision it is. This one is Revision A24. To get to this mode you will hit the 

Test button till you see “RE/A2

4

” should only take 1 or 2 clicks. 

 

 

2.

 

“REL/000” –

 This is the Relay mode. In this mode you will hit the up and 

down button to simulate gas ppm reading to set the monitor off to make 
sure they are communicating. This is one of the best troubleshooting tools 
to use for our systems.  

 

3.

 

“rFl/###” –

 This is the Signal strength mode. It allows you to see a 3 digit 

number that represents the signal strength from the head to the monitor. It 
can read from 0-205 0=no signal and 205 = best signal. This is another great 
troubleshooting tool just make sure not to leave it in this mode as it wears 
out the batteries and it jams the monitor with constant messages.  

Отзывы: