ABOUT DISINFECTION

   DISINFECTION OF THE FOLDING ARM/SELF-SUPPORTING ARM

   DISINFECTION OF THE CASE

I

   Disinfection of suction nozzle: spray disinfection water,  alcohol or other liquid  with 

disinfection function to disinfect the surface.

II  

 Disinfection of the Self – supporting arm, after daily use, please turn on the machine con      

tinuously for 30 second and then turn it off. At this time, the droplet, liquid, and dust on the 

support arm can be minimized to the greatest extent.

III

   When disinfecting regularly, please hold the support arm by hands, hold down the 

connection between the support arm and the machine, and pull the support arm out of the 

vertical direction of the device. Afterward, twist the connection of the support arm clockwise 

and disassemble it.

Separate each section of the support arm, sterilize the inner tube wall, and then assemble it 

after more than 2 hours.

IV

    Disinfect the connection between the support arm and the machine; note that after 

spraying the disinfection liquid to the connecting part, Wipe it as soon as possible and wait for 

1 hour.

  Disinfection of filter element:

The filter element is recommended to be replaced regularly.

It is strictly forbidden to wash and sterilize the filter element other than the primary effect 

water filter element and then use it again.

II 

  About the disinfection inside the chassis:

Take out all the filter element, and then spray disinfection water or other disinfection supplies 

into chassis.

After spraying , use a soft fabric to clean the inside, and just leave the case for more than 2 

hours without closing the case

.

III

   Regarding the disinfection of the disinfection layer: 

The disinfection layer needs to open the bottom cover, usually the disinfection compartment 

does not need to be disinfected, and a simple disinfection process can be performed when 

replacing the UV lamp.

#It is forbidden to use organic solvents to sterilize the equipment.

During the sterilization process, please bring a mask to prevent dust.

Содержание Aspiration

Страница 1: ...lproducts com info redlanddentalproducts com AEROSOL SUCTION Redland Aspiration Thank you for purchasing our aerosol suction machine Please read this manual carefully before use and keep it properly for future refrence ...

Страница 2: ... com info redlanddentalproducts com CONTENT OVERVIEW PRODUCT FEATURES PACKING LIST FILTERS INSTALLATION INSTRUCTIONS ABOUT DISINFECTION CONTROL PANEL AND REMOTE CONTROL FAQ AND SOLUTIONS STRUCTURE WARRANTY 1 2 3 4 5 6 7 8 9 10 ...

Страница 3: ...ilter out 99 97 of the aerosol and odor through our highly efficient customized H14 HEPA highly efficient filter and an active carbon filter The last step purifies through the UVC photo oxygen catalytic sterilization module which sterilizes and exhausts air to the outside the result our tower like filters to purify the air In conclusion this product can make the air clean and fresh to improve the ...

Страница 4: ...ses 2 4G wireless technology to achieve 360 degree control within 20 meters Easy and convient operations enable flexibility There are four levels layers of filter systems in the Redland Aspiration The systems have a washable primary effective filter water vapor absorbing filter H14 HEPA filter and an UVC photo oxygen catalytic sterilization system The Redland Aspiration ads an active carbon filter...

Страница 5: ...4 meters Folding arm User s manual 1 book Washable primary effective filter 2 piece L360xW220xH20mm AS01 H14 HEPA 1 piece L360xW220xH100mm AS02 Water vapor absorbing filter 1 piece L360xW220xH50mm AS03 Active carbon filter 1 piece L360xW220xH50mm AS04 Redland Aspiration UVC lamp 1 piece 10w AS05 Plasma 1 piece AS06 Options 2 4G Remote control 1 piece Foot Switch 1 piece AS07 Options Note The filte...

Страница 6: ...ription Replacement period Replacing method 1 Washable primary effective filter 15 30 days Pull out 2 Water vapor absorbing filter 2 4 months Pull out 3 H14 HEPA 6 8 months Pull out 4 Active carbon filter 6 8 months Pull out 5 UVC lamp 6 8 months Remove the chasis bottom cover ...

Страница 7: ...cts com 5 INSTALLATION INSTRUCTION 1 Insert one end of the power cord to the tern switch and the other end to an electrical outlet 2 Switch the machine on Note Try to make the nozzle face the mouth while using the machine to improve the suction efficiency ...

Страница 8: ...more than 2 hours IV Disinfect the connection between the support arm and the machine note that after spraying the disinfection liquid to the connecting part Wipe it as soon as possible and wait for 1 hour I Disinfection of filter element The filter element is recommended to be replaced regularly It is strictly forbidden to wash and sterilize the filter element other than the primary effect water ...

Страница 9: ...xed time with a maximum of 300 hours And minimum of 1 hour After setting press key to query the remaining time on the timer The display screen display HXXX For Example HOO5 represent the machine will auto matically shut down after hours Set the time H000 to eliminate the timer When the machine is running press the key to display HXXX on the display screen and then adjust the specific time of timin...

Страница 10: ...e machine starts it will shut down automatically repeat the above phenome non appear when turning on again Cause The internal airflow is blocked and the filter element resistance is to large which leads to excessive starting current of the fan and the machine will automatically protect and shut down Check turn off the power supply and check whether there is any foreign matter block ing the air int...

Страница 11: ...www redlanddentalproducts com info redlanddentalproducts com 9 STRUCTURE 1 Air Inlet 2 Control panel 3 Lifting handle 4 Filter s door 5 Three in one switch 6 Filters door latch plate 7 Universal wheel ...

Отзывы: