![red lion PAXDR Series Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/red-lion/paxdr-series/paxdr-series_data-sheet-manual_1427876011.webp)
11
6.1 MODULE 1 - R
ATE
S
ETUP
P
ARAMETERS
(
)
Low Update
Time
High Update
Time
Rate x
Linearizer
Segments
Rate C Display
Multiplier
Rate C
Display
Indicator
Rate x
Decimal
Position
Rate x
Scaling Input
Rate x
Scaling Display
PAR
1-rAtE
LO-Udt
C SCLr
C LIt
x
dPt
x
SE6S
x
dSP
#
x
INP
#
Pro
HI-Udt
Rate x
Display
Rounding
x
rnd
C dPt
Rate C
Decimal
Position
Rate C
Calculation
C rAtE
PARAMETER MENU
The Low Update Time is the minimum amount of time between display
updates for the Rate display. Values of 0.1 and 0.2 seconds will update the
display correctly but may cause the display to appear unsteady. The factory
setting of 1.0 will update the display every second minimum.
Note: Update time settings apply to both Rate A and Rate B.
LOW UPDATE TIME (DISPLAY UPDATE)
Module 1 is the programming for the Rate parameters. The Rate A signal is
applied to Input A, and the Rate B signal is applied to Input B. The Rate values
are shown with an indicator of
,
or
in display digit 6.
to
seconds
The High Update Time is the maximum amount of time before the Rate
display is forced to zero. (For more explanation, refer to Input Frequency
Calculation.) The High Update Time
must
be higher than the Low Update Time
and higher than the desired slowest readable speed (one divided by pulses per
second). The factory setting of 2.0, will force the display to zero for speeds
below 0.5 Hz or a pulse every 2 seconds.
HIGH UPDATE TIME (DISPLAY ZERO)
to
seconds
x = A or B Rate, # = Scaling Point (0-9)
Select the calculation for Rate C display. Rate C calculation should be set to
when not in use. When set to
, the remaining Rate C parameters are
not accessible.
SELECTION
DESCRIPTION
No Calculation. Rate C disabled.
SUM (A+B). Rate C shows the sum of Rate A and Rate B.
DIFFERENCE (A-B). Rate C shows the difference of Rate A and
Rate B.
RATIO (A/B). Rate C shows the percentage of Rate A to Rate B.
PERCENT OF TOTAL (A/A+B). Rate C shows the percentage of
Rate A to the total of Rate A and Rate B.
PERCENT DRAW (A-B/B). Rate C shows the percent draw
between Rate A and Rate B.
RATE C CALCULATION
Set the Display Multiplier to obtain the desired Rate C display resolution.
For Rate C percentage calculations, the result is internally multiplied by 100 to
show percent as a whole number. By using a Display Multiplier of 10, 100 or
1000, along with the proper decimal point position, percentage can be shown in
tenths, hundredths or thousandths respectively.
RATE C DISPLAY MULTIPLIER
Select the decimal point position for Rate C display and any setpoint value
assigned to Rate C.
RATE C DECIMAL POSITION
The Rate C display indicator in digit 6 may be turned OFF if desired. Rate A
and Rate B display indicators are always shown.
Note: Disabling Rate C indicator does NOT provide an additional digit for Rate
C value display. The display range remains -9999 to 99999.
RATE C DISPLAY INDICATOR
When Rate C is enabled, this parameter sets the decimal point position for
Rate A and B display (same position for both), and any setpoint value assigned
to Rate A or Rate B. If Rate C is disabled, this parameter does not appear and
separate decimal point positions can be selected for Rate A and Rate B.
RATE A & B DECIMAL POSITION
RATE A LINEARIZER SEGMENTS
This selects the decimal point position for Rate A display and any setpoint
value assigned to Rate A. This parameter does not appear if Rate C is enabled.
RATE A DECIMAL POSITION
This parameter specifies the number of linear segments used for the Rate A
Scaling function. Each linear segment has two scaling points which define the
upper and lower endpoints of the segment. The number of segments used
depends on the linearity of the process and the display accuracy required as
described below.
Linear Application – 2 Scaling Points
Linear processes use a single segment (two scaling points) to provide a linear
Rate display from 0 up to the maximum input frequency. For typical zero based
frequency measurements (0 Hz = 0 on display), leave
(factory setting).
For non-zero based 2 scaling point applications, set
, to enter both the
zero segment (
&
) and segment 1 (
&
).
to
Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com