Temperature 

Scale

Display 

Decimal Point

Filter Setting

User Input 

Assignment

RTD Type

SEL

tYPE

SCALE

dECPt

OFSEt

FILtr

bANd

U-ASN

1-INP

Pro

User Input 

Function

Display Offset 

Value

USrIN

Filter Band

7

6.1  MODULE 1 - S

IGNAL

 I

NPUT

 P

ARAMETERS

 (

1-INP

)

PARAMETER MENU

Pt385

tYPE

RTD TYPE

0

dECPt

0.0

0

DISPLAY DECIMAL POINT

Select the decimal point location for the desired display resolution. This 

selection applies for the Input, MAX and MIN displays.

1

FILtr

FILTER SETTING

If the displayed temperature is difficult to read due to small process 

variations or noise, increased levels of filtering will help to stabilize the display. 
Software filtering effectively combines a fraction of the current input reading 
with a fraction of the previous displayed reading to generate the new display.

Filter values represent no filtering (0), up to heavy filtering (3). A value of 1 

for the filter uses 1/4 of the new input and 3/4 of the previous display to 
generate the new display. A filter value of 2 uses 1/8 new and 7/8 previous. A 
filter value of 3 uses 1/16 new and 15/16 previous.

0,1 2 3

10

bANd

FILTER BAND

The filter will adapt to variations in the input signal. When the variation 

exceeds the input filter band value, the filter disengages. When the variation 
becomes less than the band value, the filter engages again. This allows for a 
stable readout, but permits the display to settle rapidly after a large process 
change. The value of the band is in display units, independent of the Display 
Decimal Point position. A band setting of ‘0’ keeps the filter permanently 
engaged at the filter level selected in the previous parameter.

00

   

to

  

199

 

display units

Select the RTD type used for the application. The appropriate curve will be 

automatically loaded for the selected type. The position of the Input Range 
Jumper must match the RTD type selected.

°F

SCALE

TEMPERATURE SCALE

Select the temperature scale. This selection applies for the Input, MAX and 

MIN displays.

°C

°F

RANGE JUMPERS

            TYPE

SELECTION

10 ohm

RTD Copper 10 

Cu427

100 ohm

RTD Nickel 672

Ni672

100 ohm

RTD Platinum 392

Pt392

100 ohm

RTD Platinum 385

Pt385

0

OFSEt

DISPLAY OFFSET VALUE

The temperature display can be corrected with an offset value. This can be 

used to compensate for probe errors, errors due to variances in probe placement 
or adjusting the readout to a reference thermometer.

-19999

   

to

  

19999

USER INPUT FUNCTION

NO

USrIN

Setpoint 1 and 2 Reset

Setpoint 1 Reset

Setpoint 2 Reset

Print and Reset

Reset both setpoint 1 and 2 outputs.

Resets setpoint 1 output.

Resets setpoint 2 output.

Same as Print Request followed by a 
momentary reset of the assigned value(s).

rSt12

rSt-1
rSt-2

P-r5t

Print Request

Serial transmit of the active parameters 
selected in the Print Options menu 
(Module 5).

Print

Select the value(s) to which the User Input Function is assigned. The User 

Input Assignment only applies if a selection of reset, display hold, or print and 
reset is selected in the User Input Function menu.

USER INPUT ASSIGNMENT

dSP

U-ASN

dSP

HI-LO 

LO

HI

 MODE

Reset (Edge triggered)

Program Mode Lock-out

No Function

DESCRIPTION

DISPLAY

Display Select 
(Edge Triggered)

Advance once for each activation.

d-SEL

Backlight Color 
(Edge Triggered)

Change backlight color with each 
activation (backlight version only).

COLOr

Display Intensity Level 
(Edge Triggered)

Increase intensity one level for each 
activation (backlight version only).

d-LEV

Display Hold

Holds the assigned display, but all other 
meter functions continue as long as 
activated (maintained action).

Resets the assigned value(s) to the 
current input value.

rESEt

d-HLd

See Programming Mode Access chart 
(Module 3).

P-Loc

User Input disabled.

NO

Содержание CUB5RT

Страница 1: ...this literature or on equipment must be observed to ensure personal safety and to prevent damage to either the instrument or equipment connected to it If equipment is used in a manner not specified by...

Страница 2: ...es for additional information 8 MEMORY Nonvolatile E2PROM memory retains all programming parameters and max min values when power is removed 9 CONNECTIONS Wire clamping screw terminals Wire Strip Leng...

Страница 3: ...Insert the unit into the panel cutout While holding the unit in place push the panel latch over the rear of the unit so that the tabs of the panel latch engage in the slots on the case The panel latch...

Страница 4: ...MI suppression devices or equivalent are recommended Fair Rite part number 0443167251 RLC part number FCOR0000 Line Filters for input power cables Schaffner FN2010 1 07 Red Lion Controls LFIL0000 6 To...

Страница 5: ...NNECTIONS OSNK1 OSNK2 COM 3 WIRE RTD COMM INP EXC 2 WIRE RTD COMM INP EXC JUMPER CAUTION Power input common and sensor input common are NOT isolated from user input common In order to preserve the saf...

Страница 6: ...ges to memory and return the meter to the Display Mode If power loss occurs before returning to the Display Mode verify recent parameter changes PROGRAMMING TIPS It is recommended to start with Module...

Страница 7: ...be automatically loaded for the selected type The position of the Input Range Jumper must match the RTD type selected F SCALE TEMPERATURE SCALE Select the temperature scale This selection applies for...

Страница 8: ...ess the SEL button Display reads 00r 5 Apply a direct short to terminals INP EXC and COMM using a three wire link Press SEL Display reads CALC for about 15 seconds 6 When the display reads 15 0r apply...

Страница 9: ...rogramming Immediate Access Entering Code 222 allows access regardless of security code 6 3 MODULE 3 DISPLAY AND FRONT PANEL BUTTON PARAMETERS 3 dSP SEL Pro Front Panel Display Select Enable Display U...

Страница 10: ...Balanced Hys LO bL OUTPUT STATE OFF ON Hys SP SP Hys OFF TRIGGER POINTS High Acting Unbalanced Hys HI Ub 2 HYS n HYSTERESIS VALUE 1 to 59999 Enter desired hysteresis value See Setpoint Output Figures...

Страница 11: ...ns Meter Address Abbreviated Printing 5 SEr bAUd dAtA PAr Addr Abbr OPt Pro The Serial Setup Parameters are only active when one of the optional serial communications programming cards is installed in...

Страница 12: ...it Only Programming Security Code CodE Display Decimal Point RTD Type tYPE 1 INP Temperature Scale SCALE dECPt USrIN User Input Function Filter Setting Display Offset Value OFSEt FILtr Filter Band bAN...

Отзывы: