red lion APLTC Скачать руководство пользователя страница 1

For Sales and Support, Contact Walker EMD • Toll-free: (800) 876-4444 • Tel: (203) 426-7700 • Fax: (203) 426-7800 • www.walkeremd.com

1

!

MICROPROCESSOR CONTROLLED, NO POTS OR DIP

SWITCHES

!

USER PROGRAMMABLE T/C TYPE (T, E, J, K, R, S or B)

!

SELECTABLE °F OR °C DISPLAY

!

FULL 4-DIGIT, 0.56" (14.2 mm) RED LED READOUT

!

FAULT DETECTION FOR THERMOCOUPLE BURNOUT

!

OPTIONAL ISOLATED ANALOG OUTPUT

!

OPTIONAL ISOLATED SERIAL COMMUNICATIONS

!

OPTIONAL ISOLATED TEMPERATURE ALARM OUTPUT

!

NEMA 4/IP65 SEALED METAL FRONT BEZEL

DESCRIPTION

The Apollo Thermocouple Indicator is a high accuracy, microprocessor

based instrument designed for low cost temperature measurement and control.
The microprocessor automatically compensates for cold junction, NBS
linearity, and the meter’s zero and span. 

(One model accepts all thermocouple

types by simple programming.)

There is no need to recalibrate when changing

input types. Isolated analog output, serial communications, and temperature
alarms are available as low cost options.

The NEMA 4/IP65 designed, die-cast metal bezel of the Apollo can be sealed

in the front panel for use in wash-down and other tough industrial environments.
The full 4-digit display features large 0.56" 

(14.2 mm)

red LEDs for easy reading. 

SPECIFICATIONS

1. 

DISPLAY:

4-digit

(9999)

, 0.56"

(14.2 mm)

high LED, minus sign displayed

when temperature is negative

2. 

RESOLUTION:

1 degree

3. 

POWER:

Available in two voltage ranges 115 or 230 VAC, ±10%, 50/60 Hz,

8 VA

4. 

THERMOCOUPLE TYPES:

T, E, J, K, R, S and B

5. 

INPUT IMPEDANCE:

10 M

(protected from 115 VAC faults)

6. 

LEAD RESISTANCE EFFECT:

20 µV/350 

7. 

OPEN THERMOCOUPLE INDICATION:

Display

“OPEN”

Analog Output

: -500 mV output

Serial Output

“OPEN”

in data field

8. 

COLD JUNCTION COMPENSATION ERROR:

0.02 degree/degree

9. 

READING RATE:

2.5 readings/second

10. 

RESPONSE TIME:

2 seconds for step change

11. 

NORMAL MODE REJECTION:

40 dB at 50/60 Hz

12. 

COMMON MODE REJECTION:

110 dB, DC OR 50/60 Hz

13. 

TEMPERATURE EFFECTS:

Operating Range

: 0° to +50°C

Storage Range

: -40° to +80°C

Relative Humidity

: Less than 85% R.H.

Span Temperature Coeff.

: ±60 ppm/°C

Zero Temperature Coeff.

: ±0.3 µV/°C

14. 

SERIAL COMMUNICATIONS (optional):

20 mA current loop

Data Format

: 10 Bit Frame, Odd Parity

Baud Rates

: 300, 600, 1200, 2400

15. 

ALARMS (Optional):

(2)

isolated 

(output COMM. common with analog

output COMM.)

, solid-state, current sinking NPN open collector transistors.

Resolution

: 1 degree

Hysteresis

: programmable, min. 1 degree

Imax

: 100 mA @ V

SAT

= 1 V

Vmax

: 30 VDC 

(Zener protected)

Connections

: 3-position, removable terminal block

16. 

ANALOG OUTPUT (Optional):

1 mV/degree, bipolar, 10 mA max.

Isolated 

(analog output COMM. common with alarm output COMM.)

Accuracy

: ±6 mV

Adjustments

: zero and span pots

Connections

: 5-position, removable terminal block

17. 

CONSTRUCTION:

Die-cast metal front bezel that meets NEMA 4/IP65

requirements for wash-down and/or dusty environments when properly
installed. Case body is black high impact plastic (panel gasket and mounting
clips included).

18. 

CONNECTION:

6-position terminal block

19. 

WEIGHT:

1.2 lbs. 

(0.6 kg)

MODEL APLTC - APOLLO THERMOCOUPLE INDICATOR

DIMENSIONS  In inches (mm)

Note: Recommended minimum clearance (behind the panel) for
mounting clip installation is 2.1" (53.4) H x 5.5" (140) W.

Bulletin No.  APLTC-I
Drawing No.  LP0154
Released  2/05

Tel +1 (717) 767-6511
Fax +1 (717) 764-0839
www.redlion.net

PANEL CUT-OUT

Содержание APLTC

Страница 1: ...FECT 20 V 350 7 OPEN THERMOCOUPLE INDICATION Display OPEN Analog Output 500 mV output Serial Output OPEN in data field 8 COLD JUNCTION COMPENSATION ERROR 0 02 degree degree 9 READING RATE 2 5 readings...

Страница 2: ...rejection A non volatile E2PROM memory device provides permanent data retention for microprocessor variables The display consists of 5 solid state LEDs for temperature and units readout The serial com...

Страница 3: ...pidly The button can be released and pressed again to increment slowly when the desired value is approached SERIAL COMMUNICATIONS OPTION DESCRIPTION The Apollo Thermocouple Indicator Serial Communicat...

Страница 4: ...Apollo must perform and the identifier on which it operates 3 If the change command V is being used the next characters in the string after the value identifier are the numerical data 4 All commands...

Страница 5: ...temperature 1500 2000 etc Adjust the span potentiometer left side so that the reading from the voltmeter and the indicator correspond 3 Repeat steps 1 and 2 until both points are correct Calibration i...

Страница 6: ...anneal one type of alloy it must be brought to 700 F and left to soak for 2 hours after which it must cool at the rate of 100 hour until it reaches 500 where it must stay for 1 hour and then allowed t...

Страница 7: ...trudes through the other side Install each of the mounting clips by inserting the wide lip of the clips into the wide end of the hole located on either side of the case Then snap the clip onto the cas...

Страница 8: ...or products containing components manufactured by RLC and based upon personal injuries deaths property damage lost profits and other matters which Buyer its employees or sub contractors are or may be...