9. Troubleshooting
Note:
Before continuing, check all accessories visually for any defects.
Error No.
Error Message
Corrective Action
1
Air Leak
See Table 9
2 to 6
Internal Error
See Table 9
The following is an error message example (Air Leak):
Table 9 - Troubleshooting Guide
Symptom
Possible Cause
Corrective Action
The console is not
working.
No electricity and no
battery power.
Check the electrical wall outlet.
Additionally replace the battery
pack.
DC power adapter
Verify that the DC power adapter is
connected to the DC adapter socket on
the console, and the power cord is
connected to the 115-Volt wall outlet.
The console starts
working and stops
immediately.
The air cannot move
through the hose
bundle.
Check hose bundles for kinks, twists and
folds.
One garment inflates but
the second one does not.
The second garment
does not receive air.
Check its hose bundle for kinks, twists
and folds.
The console stops
working, Error 1 message
displayed, air leak icon
turns On, and buzzer
beeps.
Hose bundle is not
connected properly to
garment or console, or
Prong Plug is not
inserted into unused
air outlet.
Check and fasten all air connections. If
treating only one limb, make sure the
unused air outlet is plugged with the
prong plug provided with the console. If
all air connections are OK and problem
persists, contact Recovery Pump LLC.
The console works at a
very low pressure,
regardless of the pressure
set by the user.
Defective garment
Replace garment and check again.
Internal malfunction
Contact Recovery Pump LLC.
An irregular noise.
Console transferring
vibrations to a surface
Make sure the console is standing evenly
on all four of its bumpers.
Internal malfunction
Contact Recovery Pump LLC.
L98005H1R-A
29