4. If the resistance is lower than 30 Ω, the buzzer
will buzz.
Diode measurement
1.
Set the rotary switch to
position.
2. Press the SELECT button, and the icon will appear
.
3. Connect the red lead to anode of the diode, and
black to the cathode.
4. Voltage drop will be displayed. If the display
shows the "OL", it means that the polarization is
reversed.
Capacitance measurement
1.
Set the rotary switch to
position .
2. The default range is nF.
3. Connect the test leads to the capacitor. Make
sure that the polarization is correct.
Note:
• Before measurement discharge all of the
capacitors.
• During measurement of capacitors higher than 10
µF it may take a while for the result to stabilize.
Frequency and duty cycle measurement
• Set the rotary switch to the Hz% position.
• Press the Hz% button to switch between
frequency and duty cycle measurement.
• Connect the test leads to a measured circuit. The
result will appear on the display.
Caution: input voltage should be in range between
200 mV and 10 V AC. If the voltage exceeds the
EN
15
Содержание MIE-RB10B
Страница 1: ...TOOLS model MIE RB10B USER S MANUAL DE EN PL RO POCKET MULTIMETER ...
Страница 2: ...DE 2 ...
Страница 37: ...RO 37 ...
Страница 38: ......
Страница 39: ......