background image

 

 

4. 

CONTROLS 

  

 

4.01  Output Mode Link:  

 

The "OUTPUT" jumper link is set to "V" for voltage output or "mA" for current output. 
Three 20-turn screwdriver-adjusted controls are provided for controlling: 

 

4.02  Excitation  

 

- allowing continuous variation of excitation between 3V and 10V.  

Note

: if supply is 

<13V,  the  max.  excitation  voltage  available  is  Vs  -  3V. 

Note

:  for  units  of  MOD18 

onward, if operating temperature exceeds 70°C then the maximum current is 60mA, 

 

4.03  Zero  

 

-  allowing  a  fine  adjustment  of  output  zero  of  approximately  ±2V  (or  0-10mA) 
depending on Fine Gain setting. 

 

4.04  Fine Gain  
 

- allowing a 4:1 variation in amplifier gain, used in conjunction with 4.05. 

 

4.05  Gain Range (x1 to X1250) 

-

  sliders  1  to  4  of  the  6-slider  DIL  switch,  when  used  with  the  fine  gain  control, 

provide an output of 5V or 4-20mA with an input signal range of approx. 4mV to 
5V as shown below: 

(also output ± 5V for input ±4 mV to ±5V) 

 

SLIDERS 

ON 

For Mod.8 (and above) 

For ±10V 

OUTPUT 

INPUT SIGNAL RANGE for: 5V or 4-20mA OUTPUT 

4 + 3 + 2 

– 16mV 

±8 to ±32mV 

4 + 3 

– 20mV 

±10 to ±40mV 

4 + 2 

18 

– 70mV 

±35 to ±140mV 

4 + 1 

65 

– 270mV 

±130 to ±500mV 

0.25 

– 1V 

±0.5 to ±2V 

0.34 

– 1.35V 

±0.7 to ±2.7V 

None 

1.25 

– 5V 

±2.5 to ±10V 

Note:  for the two lowest gain ranges (with slider 4 off), the excitation voltage must be 

5V 

to provide signals within the required common mode voltage range of the amplifier. 

Note:  for ±10V output, the supply must be at least 22V or ±11V (see 4.10). 
 

4.06  Shunt Calibration  
 

- slider 5 ("C") of the 6-slider DIL switch, when set to ON connects an internal 59k

Ω

 

0.1% resistor across one arm of the bridge. 

 
Remote  Cal. 

-  an  internal  relay  may  be  used  to  remotely  connect  the  59k

Ω

  resistor.  To 

operate this relay, connect the R-Cal. terminal to V- (CON 2) at any convenient point (lead 
length is irrelevant) via a switch, etc. 
 
Note  1:  For  normal  relay  operation,  the  supply  voltage  must  be  at  least  15V.    For  lower 
supply voltages, solder link SP2 must be made (see Fig.1). 
 
Note 2: For normal shunt calibration, connect the CAL terminal to Sig- (CON 1).  This links 
the internal calibration resistor (terminal 6) to the relevant bridge node.  
 
 

Содержание S7DC

Страница 1: ...Declaration of Conformity USA Canada All other countries RDP Electrosense Inc RDP Electronics Ltd 2216 Pottstown Pike Pottstown PA 19465 U S A Grove Street Heath Town Wolverhampton WV10 0PY United Ki...

Страница 2: ...4 05 Gain Range x1 to X1250 9 4 06 Shunt Calibration 9 4 07 Bandwidth Noise Capacitor 10 4 08 4 20mA Offset Solder Link SP1 10 4 09 Relay Voltage Solder Link SP2 10 4 10 Maximum Voltage output vs Supp...

Страница 3: ...ar eg 4mV input signals and can produce corresponding uni polar or bi polar voltage outputs ie 0 to 10V or 10V sensitivities and gains given in this document refer to both Current outputs 4 to 20mA ar...

Страница 4: ...clare that the product described in this technical manual is manufactured by RDP Electronics Limited and performs in conformity to the following The Electromagnetic Compatibility Directive 2014 30 EU...

Страница 5: ...y from possible sources of EMI e g motor solenoid relay cables etc 2 5 The transducer body should be grounded The transducer body being clamped or bolted to grounded metalwork e g on a machine usually...

Страница 6: ...e output dual supply supply must be symmetrical Either the supply V or the output common 0V or both must be fully floating ie only one can be grounded Failure to do this may result in damage to the am...

Страница 7: ...ion on the underside of the S7DC PCB This requires removal of the PCB from the case R22 replaces bridge arm D R23 replaces bridge arm A R24 replaces bridge arm B Note Mounting these resistors in the i...

Страница 8: ...metal outer shell of the gland The bore of the gland is a tight fit onto the cable shield giving the required ground contact 3 Fit gland cap and tighten Cable Cores Trim Rubber Seal Cable Shield Doub...

Страница 9: ...For 10V OUTPUT INPUT SIGNAL RANGE for 5V or 4 20mA OUTPUT 4 3 2 4 16mV 8 to 32mV 4 3 5 20mV 10 to 40mV 4 2 18 70mV 35 to 140mV 4 1 65 270mV 130 to 500mV 4 0 25 1V 0 5 to 2V 1 0 34 1 35V 0 7 to 2 7V N...

Страница 10: ...e g for compressive tensile load applications SP1 may be fitted which changes the relationship to full scale input signal 2 5V or 4 20mA output i e SP1 provides a F S offset for the current output an...

Страница 11: ...ontrol for zero output 5 4 Apply a precise load pressure etc to the transducer and adjust the F GAIN control for the desired output 5 5 Repeat steps 3 and 4 until repeatable readings are obtained 5 6...

Страница 12: ...re or load to the transducer 7 2 1 Calculate the shunt calibration figure required from the data given on the transducer Calibration Record Sheet CRS From CRS output for 100 WmV output with shunt YmV...

Страница 13: ...load change 0 100 mA Tempco 0 005 C typical Amplifier Linearity 0 02 of FS output Bandwidth 5kHz typical See Section 4 07 20Hz with filter ON Noise 10mV or 30 A p p less with filter ON typical Zero Te...

Страница 14: ...itional error can occur at certain frequencies Field Strength Typical Maximum Error 10V m 5 3V m 0 5 Mech Env mental Working Temp Range 10 to 60 C For units of MOD18 onwards the temperature range is 4...

Страница 15: ...gether and repeat B E Check the transducer is correctly wired F Check that SUPPLY and OUTPUT circuits are wired correctly and not simultaneously grounded if using SINGLE SUPPLY G Disconnect transducer...

Страница 16: ...repairs and investigations are subject to at least the minimum charge prevailing at the date of return The type and serial number of the instrument should always be quoted together with full details o...

Отзывы: