CONTOUR® NEXT ONE
Glucose Meter
PAGE 30
REMOTE CARE PARTNERS
V1.0
Control Solution Testing
1.
Remove a Contour next test strip from the
bottle.
2.
Tightly close the bottle lid immediately after
you remove the test strip.
3.
Hold the test strip with the grey square end
facing up.
4.
Insert the grey
square end of the test
strip into the test strip
port until the meter
beeps. The meter
turns on, displaying a
test strip with a
flashing blood drop.
NOTE: If this is the first time you are opening the
control solution, write the date on the bottle
5.
Shake the control solution bottle
well, about 15 times before every use.
6.
Remove the bottle cap and use a tissue to
wipe away any solution around the bottle tip
before dispensing a drop.
CAUTION:
Do not use control solution that is more
that 6 months past the date you first opened the bottle.