background image

RBR#0008819revA

 –  

6

5  RBRquartz BPR|zero

5.1  Overview - RBRquartz BPR|zero

5.1.1  Description

The RBR

quartz

³ BPR | zero is a special version of the robust RBR

quartz

³ BPR implementing an internal quartz barometer 

and a switching valve. The novel AzeroA technique is used to provide in-situ reference measurements to correct for long 
term drift in the Paroscientific Digiquartz® pressure gauge. Configured with one or two Digiquartz® pressure gauges, an 
internal barometer is used in conjunction with a hydraulic manifold to periodically make reference measurements of 
internal housing pressure. This instrument is intended for deep water, long-term deployments where high stability and 
resolution of absolute pressure measurements are critical. The RBR

quartz

³ BPR | zero may be supplied with external 

power from RBR

fermata

 (power) or RBR

cervata

 (power and memory) canisters or cabled to an observatory for external 

power and realtime data access.

The RBR

quartz

³ BPR | zero uses the AzeroA technique to assess drift in the Digiquartz® pressure gauge. This is done by 

periodically switching the applied pressure that the gauge measures from seawater to the atmospheric conditions 
inside the housing. The drift in quartz sensors is proportional to the full-scale rating, so a reference barometer - with 
hundreds of times less drift that the marine gauge - is used to determine the behaviour of the marine pressure 
measurements.

5.2  Specifications - RBRquartz BPR|zero

Physical

Weight:

~30kg in air (one Paros)

~18kg in water (one Paros)

Size:

~750mm x Ø140mm

Housing:

Titanium

Depth rating:

7000 m

Operational temperature:

-20°C to 50°C*

Storage temperature:

-20°C to 50°C

Содержание quartz3 BPR zero

Страница 1: ...rbr global com INSTRUMENT GUIDE BPR zero...

Страница 2: ...9 6 2 Changing the batteries RBRquartz BPR zero 11 6 2 1 Steps 11 6 3 Installing desiccant RBRquartz BPR zero 12 6 4 Maintenance and repairs 12 6 4 1 O rings RBRquartz BPR zero 12 Inspecting O rings 1...

Страница 3: ...0008819revA 3 2 Introduction You can access the RBR instrument guide on the USB data stick provided when you purchase a logger from the Help menu in Ruskin and on the RBR web site at www rbr global c...

Страница 4: ...RBR 0008819revA 4 3 Revision history Revision No Release Date Notes A 04 May 2020 Original...

Страница 5: ...by us The liability of RBR Ltd extends only to the replacement cost of the instrument The customer will bear all costs of shipment to us for repair all other costs including return shipment will be b...

Страница 6: ...rements are critical The RBRquartz BPR zero may be supplied with external power from RBRfermata power or RBRcervata power and memory canisters or cabled to an observatory for external power and realti...

Страница 7: ...35 C The instrument has 8 X AA batteries installed however this is intended only for purposes of sustaining the instrument during external power interruptions of a few minutes While the Paros will con...

Страница 8: ...2000 4000 7000 dbar Initial accuracy 0 01 FS full scale Resolution 10ppb at 16Hz sampling rate Accelerometer optional Range 3g Resolution 100ng The full mechanical drawing is available in Appendix A...

Страница 9: ...shown below Only the battery end cap needs to be opened to access the data connector batteries desiccant and waste reservoir To open the logger battery end cap the two M6 screws will need to be remov...

Страница 10: ...il the battery end cap is mated fully to the housing Remember to keep the O rings clean and avoid scratching the O ring mating surfaces Carefully inspect the O rings as described in the Inspecting the...

Страница 11: ...ent during external power interruptions of a few minutes While the Paros will continue to run and be measured the valve will not operate 6 2 1 Steps Remove the battery end cap from the logger see Open...

Страница 12: ...desiccant capsules can be opened the desiccant removed and replaced or dried 6 4 Maintenance and repairs 6 4 1 O rings RBRquartz BPR zero The single most important item of maintenance on any RBR subme...

Страница 13: ...bishment please contact us for advice Replacing an O ring Lever the O ring from its groove Use a soft plastic or wood tool do not use a metal screwdriver or any other tool which may scratch the surfac...

Страница 14: ...easure the pressure inside the housing of the instrument which is monitored by a barometer in order to correct for the drift of the sensors over time This actuation of this valve results in a small vo...

Страница 15: ...towel around the reservoir in a clockwise manner when viewed as below The towel should only wrap around the reservoir once any additional layers will make it difficult to insert 8 While holding the to...

Страница 16: ...closely as possible to the current market value based on the original purchase price less depreciation Please contact RBR if you have any questions about the instrument value Send the unit using your...

Страница 17: ...RBR 0008819revA 17 7 Appendix 7 1 Appendix A RBRquartzBPR zero Product Drawing...

Страница 18: ...RBR 0008819revA 18...

Страница 19: ...RBR 0008819revA 19 7 2 Appendix B RBRquartzBPR zero Service Drawing...

Страница 20: ...RBR 0008819revA 20 7 3 Appendix C RBRquartzBPR zero Connector Details...

Отзывы: