RBR#0008817revC
6
PAR and narrow-band radiometers
The RBR
coda
3
PAR and RBR
coda
3
rad instruments look identical and share several specifications.
The RBR
coda
3
PAR and RBR
coda
3
PAR
|
deep use the cosine photosynthetically active radiation sensors.
The RBR
coda
3
rad and RBR
coda
3
rad
|
deep use radiometers measuring narrow-band light with a fixed channel width,
available in various 10nm- and 25nm-wide channels. Both centre wavelength and channel width are factory-
configured.
Optical radiometry
Parameter
Value
Initial offset error*
±0.0025% full scale
Resolution**
±0.0002% full scale
Dynamic range
>5.5 decades (nominal)
Absolute calibration***
±5%
Linearity
±1%
Time constant
<5ms
Operating temperature
range
-5°C to 35°C
Gain temperature
dependence
0.15% / °C
Cosine response error
(water)
±5% at 0-60°,
±10% at 61-82°
Azimuth error (water)
±1.5% at 45°
Out-of-band rejection**
>25dB (typical), OD 2.5
* Dark offset is internally temperature-compensated.
** Out-of-band rejection and resolution are wavelength
dependent for narrow-band radiometers.
*** RBR calibrates radiometers with NIST traceable
references.
PAR
Parameter
Value
Wavelength range
400 to 700nm
Full scale range
0 to 5000µmol/m
2
/s (minimum)
Initial offset error*
±0.125umol/m²/s
Resolution
±0.010umol/m²/s
.
Narrow-band channels
Parameter
Value
Centre
wavelengths (CWL)
413/ 445/ 475/ 488/ 508/ 532/
560nm
Accuracy
(for CWL)
±3nm (for all CWLs except 475nm)
±5nm (for CWL 475nm only)
Full width at half-
maximum (FWHM)
10nm (for all CWLs except 475nm)
25nm (for CWL 475nm only)
Accuracy
(for FWHM)
±3nm
Full scale range
0 to 400µW/cm
2
/nm (minimum)
Initial offset error*
±0.010µW/cm²/nm
Resolution**
±0.001µW/cm²/nm