ASCII Programming
112
MI3
Description
Char
Format
P B S N Legal Values
Factory default Head /
Box
Profibus / Modbus
Address
XAS
n
√
√
√
0 – 125 (Profibus)
1 – 247(Modbus)
0 (Profibus)
1(Modbus)
B
Device Bottom range
limit
XB
?nXB
float
√
√
answer !nXB=10.0
H
Restore Box Factory
defaults
XF
√
except address B
Transmission
XG
float
√
√
√
0.1 – 1.0
1.0
H
Device High range limit XH
?nXH
float
√
√
read only
H
Sensor initialization
XI
1 after reset
0 if XI=0
1
B
Box Temperature
XJ
float
(
°
C/
°
F)
B
FTC3 trigger/hold
XN
nXN
T = trigger, H = hold
T
H
analog Output 1 mode XO1O
(XO)
XO1O=v
(XO=v)
√
√
0 – 0...20 mA
4 – 4...20 mA
5 –
TCJ
6 – TCK
7– TCR
8 – TCS
9 – 0...5 V
10 – 0...10 V
99 – disable (tristate)
XO1O=9
for
communication
box (metal)
XO1O=9
9 for
communication
box 6TE, analog
B
analog Output 2 mode XO2O
XO2O=v
√
√
0 – 0...20 mA
4 – 4...20 mA
9 – 0...5 V
10 – 0...10 V
99 – disable (tristate)
XO2
O=
4 for
communication
box (metal)
XO1O=9
9 for
communication
box 6TE, analog
B
analog Output 3 mode XO3O
XO3
O=v
√
√
0 – 0...20 mA
4 – 4...20 mA
9 – 0...5 V
10 – 0...10 V
99 – disable (tristate)
XO3
O=9
9 for
communication
box 6TE, analog
analog Output 4 mode XO4O
XO4
O=v
√
√
0 – 0...20 mA
4 – 4...20 mA
9 – 0...5 V
10 – 0...10 V
99 – disable (tristate)
XO4
O=9
9 for
communication
box 6TE, analog
Box Firmware Revision XR
?XR
√
e.g. 1.01
Set in FW
B
Setpoint relay function XS
nXS
float
√
√
°C/°F
500°C
H
Trigger
XT
?XT
√
√
√
0 = inactive, 1 = active
0
B
Unit identification
XU
?XU
√
e.g. !XUMI3COMM
set at production
Serial number
XV
?XV
√
e.g. !XV98123
set at production B
Advanced hold
hysterese
XY
nnnn
√
√
H
1
Mode 0 and 4 for MI3MCOMMA only
2
Mode 5 to 8 for MI3COMM only
1.800.868.7495
Содержание MI3
Страница 2: ...1 800 868 7495 info Raytek Direct ca...
Страница 9: ...20 NOTES 119 1 800 868 7495 info Raytek Direct ca...
Страница 73: ...Accessories MI3 Rev F1 05 2013 73 Figure 66 Protective Window 1 800 868 7495 info Raytek Direct ca...
Страница 113: ...ASCII Programming MI3 Rev F1 05 2013 113 1 800 868 7495 info Raytek Direct ca...
Страница 119: ...Notes MI3 Rev F1 05 2013 119 20 Notes 1 800 868 7495 info Raytek Direct ca...