Installation
16
MI
5.2.2
Cable
for
Power
Supply
and
Outputs
You
need
to
connect
the
power
supply
(12
to
26
VDC)
and
the
signal
output
wires.
Use
only
cable
with
outside
diameter
from
4
to
6
mm
(0.16
to
0.24
in),
AWG
24.
The
cable
must
include
shielded
wires.
It
should
not
be
used
as
a
strain
relief!
1.
Cut
about
40
mm
(1.5
in)
of
the
cable
sheath
(7)
from
the
end
of
the
cable.
Caution:
Do
not
cut
into
the
shield!
2.
Cut
the
shield
(5)
so
about
5
mm
(0.2
in)
remains
exposed
from
under
the
cable
sheath
(7).
Separate
the
shield
and
spread
the
strands
out.
Shorten
the
inside
insulation
until
you
can
separate
the
wires
(6).
3.
Strip
3
mm
(0.15
in)
of
insulation
from
the
wires
(6).
Figure
5:
Cable
Preparation
4.
Open
the
electronic
box
by
removing
the
four
Phillips
head
screws
and
pulling
off
the
lid.
Unscrew
the
cap
(1),
and
remove
the
plastic
compression
fitting
(2),
the
rubber
washer
(3),
which
is
inside
the
fitting,
and
the
two
metal
washers
(4).
Содержание MI Miniature Infrared Sensor
Страница 1: ...MI Miniature Infrared Sensor Operating Instructions Rev F 04 2006 54301...
Страница 56: ...Accessories 48 MI 8 2 Adjustable Mounting Bracket Figure 31 Adjustable Mounting Bracket XXXMIACAB...
Страница 57: ...Accessories MI 49 8 3 Fixed Mounting Bracket Figure 32 Fixed Mounting Bracket XXXMIACFB...
Страница 63: ...Accessories MI 55 Figure 39 Air Cooling System T Adapter...
Страница 64: ...Accessories 56 MI Figure 40 Dimensions of Air Cooling System Hose inner 9 mm 0 35 in outer 12 mm 0 47 in...
Страница 66: ...Accessories 58 MI 8 7 Box Lid Figure 43 Box Lid with View Port for Post Installations XXXMIACV...