Technical Data
22
Marathon MM
3.3
Electrical Specifications
Power Supply
24 VDC ± 20%, min. 500 mA
Outputs
Analog
0 - 20 mA, 4 - 20 mA (active)
14 bit resolution
max. current loop impedance: 500
Ω
RS485 Interface
networkable to 32 sensors
Baud rate: 300, 1200, 2400, 9600, 19200, 38400 (default), 57600,
115200
(max. 38400 Baud in 2-wire mode)
Data format: 8 bit, no parity, 1 stop bit,
4-wire mode (full-duplex) or 2-wire mode (half duplex),
selectable via control panel or software
Relay
Contacts max. 48 V, 300 mA, response time < 2 ms,
(software programmable)
Display
5 digit backlit LCD display
External Input
Input Voltage
0 to 5 VDC
functions: trigger, laser ON/OFF switching, ambient
background temperature compensation, or emissivity setting,
see section 7.4.3
page 44.
3.4
Environmental Specifications
Environmental
rating
NEMA-4 / IEC 529, IP 65 (also with video option)
Relative Humidity
10% to 95% non-condensing
Storage Temperature
-20 to 70°C (-4 to 158°F)
Ambient Temperature
5 to 65°C
(41 to 149°F) without cooling
with video
5 to 50°C (41 to 122°F) without cooling
with air cooling
10 to 120°C (50 to 250°F)
with water cooling
10 to 175°C (50 to 350°F)
with ThermoJacket
10 to 315°C (50 to 600°F) water cooled
Warm up Period
20 min.
Vibration
MIL-STD-810D (IEC 68-2-6) 2 G, 10 - 150 Hz, 3 axis
Mechanical Shock
MIL-STD-810D (IEC 68-2-27) 5 G, 11 ms duration, 3 axis
Weight
0.7 kg (1.54 lb)