![RaySafe i2 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/raysafe/i2/i2_installation-and-maintenance-manual_3941578021.webp)
21
RaySafe i2 Installation and maintenance manual – Installation
ConfIgURe tHe doSImeteR In doSe mAnAgeR oR doSe vIeweR
Figure 13. The dosimeter options dialog box
follow the steps below to configure a dosimeter:
1. Put the dosimeter in the cradle. within a few seconds dose viewer/manager will have detected the
dosimeter.
2. dose view/manager will start loading the dosimeter dose history. follow the progress in the progress bar in
the lower left corner.
3. open the dosimeter options dialog box. Press the Settings tab and make the following adjustments:
• Set the dosimeter’s Power mode to on. this starts the radio communication and the dosimeter is ready
to start logging dose data.
• Set the real time display mode to Show, which makes the dosimeter appear on the real time displays.
• verify that the windows time settings on the PC are correct. Synchronize the dosimeter’s clock with the
computer’s clock. (be sure to do this for all dosimeters in one lab to get them synchronized.)
• Select Reset dose History.
• verify that battery Status is green or orange.