@2021 Raycus Corporation 

 

 

 

f)

 

The  device  has  four  cooling  fans  at  the  rear  panel  to  dissipate  heat.  In  order  to 

guarantee enough  airflow to  help  giving heat  off, there must be  a space  of at  least  10 

cm’s  width  for  airflow  in  front  and  rear  side  of  the  device.  As  the  cooling  fans  are 

working  at  blow  condition,  if  laser  is  mounted  in  a  cabinet  with  fans,  the  direction 

should be same as laser’s fans. 

g)

 

Do not look into the output head of the device directly. Please do wear appropriate laser 

safety eyewear during operation of the device. 

h)

 

The longest time without pulse is only 100

s

. If there is no pulse output, please stop 

marking at once to avoid the further damage of the device. 

i)

 

Power source sudden interruption will do  great  harm to  the laser device.  Please make 

sure the power supply works continuously. 

1.4.

 

Specifications 

Table 2 P100MX pulsed fiber laser specifications 

 

NO. 

Characteristic 

Test condition 

Min 

Typ. 

Max. 

Unit 

Mode of operation 

 

Pulsed

 

 

Polarization 

 

Random 

 

Maximum pulse energy 

350ns/66kHz/100W 

1.5 

mJ 

Nominal average output power 

 

100 

103 

105 

Pulse duration 

Subject to Table 3

 

10 

350 

ns 

Output power adjustment 

range 

 

100 

Long-term average power 

instability 

 

 

 

Pulse repetition rate 

Subject to Table 3

 

2000 

kHz 

Central emission wavelength 

FWHM 

1059 

1064 

1069 

nm 

10 

Emission bandwidth 

FWHM 

10 

nm 

11 

Beam quality M

2

 

350ns/66kHz/100W

 

1.3 

1.6 

 

12 

Output beam diameter 

350ns/66kHz/100W 

mm 

13 

Laser switching ON time 

0-100%  power 

100 

μ

14 

Laser switching OFF time 

100-10%  power 

30 

μ

15 

Output fiber cable length 

 

16 

Supply voltage 

 

23 

24 

25 

VDC 

17 

Guide Laser Power 

 

0.2 

0.5 

mW 

18 

Anti Strong Back Reflection 

 

Yes 

 

19 

Cooling 

 

Forced Air Cooled 

 

20 

Operating Temperature 

 

40 

 

Содержание RFL-P100MX

Страница 1: ...A0 S RFL P100MX 100MX RFL P100MX JS02 20210315 20210316 20210322 20210322 20210322...

Страница 2: ...User Instruction RFL P100MX Wuhan Raycus Fiber Laser Technologies CO Ltd 2021...

Страница 3: ...make sure that the correct voltage of 24V DC is connected and the anode and cathode are right Incorrect polarity or voltage will harm severely the device b The device emits invisible light with a wav...

Страница 4: ...ions 1 1 4 Specifications 2 2 Mounting 4 2 1 Mounting dimensions 4 2 2 Method of installation 4 3 Control Interface 6 4 Operation Regulations 12 4 1 Pre inspection 12 4 2 Operation procedures 12 4 3 C...

Страница 5: ...ory inspection report 1 1 3 Environmental requirements and cautions Pulsed laser should be driven by 24V DC 1V power source a Caution Make sure the corresponding wires of the device are properly groun...

Страница 6: ...make sure the power supply works continuously 1 4 Specifications Table 2 P100MX pulsed fiber laser specifications NO Characteristic Test condition Min Typ Max Unit 1 Mode of operation Pulsed 2 Polari...

Страница 7: ...5W 2 20 1 0 2 4W 500 100 105W 2000 100 105W 3 30 1 0 3 4W 330 100 105W 1000 100 105W 4 60 1 0 5 4W 200 100 105W 1000 100 105W 5 100 1 0 8 4W 125 100 105W 1000 100 105W 6 200 1 1 25 4W 80 100 105W 500...

Страница 8: ...on drawing of output isolator Unit mm The output head of the isolator is subject to the final factory product 2 2 Method of installation a Fix the module stable to the bracket and keep the laser in go...

Страница 9: ...external controller matches the laser and the control cable is well connected to the laser s interface The recommended electrical connection is shown in Figure 5 Figure 5 Wiring diagram of recommended...

Страница 10: ...codes in Table 5 17 VCC 5VDC power supply input Providing power for inside isolate chip of DB25 to ensure that the input and output signal are valid 18 EE Emission Enable EE signal HIGH 4 5V Emission...

Страница 11: ...signal of the MO Pin19 is the input for the optical output signal The electrical level for both Pin18 and Pin19 are 5V Before turning on Pin 19 MO signal must be switched ON in other word the signal o...

Страница 12: ...factory which can meet the needs of most users Special needs shall be put forward in the customized contract g Definition of alarm signal Table 6 Definition of alarm signal Pin12 Pin11 Pin16 Pin21 Ala...

Страница 13: ...ition of Serial Communication Pin Name Description 2 Serial input The setting of the data bits synchronize with rising edge of the serial clock 3 Serial clock Serial clock frequency range 10KHz 100KHz...

Страница 14: ...4 The input data length is 4 bytes the first byte of instruction code should be 0x01 01h the second byte and the third byte are the value of the binary coding of laser duration for example 200 is the...

Страница 15: ...parameter should show up select Fiber in the category of the Laser Control panel follow that select IPG_YLPM and then close this window c Pulse duration can be modified by the Duration on the right s...

Страница 16: ...s showed in the Table 3 c Laser marking checking For the device initial testing first turn the power down to zero without turning on the marking system after the device is successfully started Then dr...

Страница 17: ...not following the instructions caused by those who are not from Raycus Customer has the responsibility to understand and follow this instruction to use the device Any damage caused by fault operating...

Отзывы: