In
case
if
not
all
marks
are
found
on
the
image,
you
will
also
see
the
error message. It
may mean the
following:
●
not
all
markers
are
visible
on
one
of
the
snapshots,
●
not
all
markers
are
lit
by
the
projector,
●
calibration
plate
is
mounted
at
large
angle
to
the
scanner,
●
plate
is
too
close
or
too
far
from
the
scanner
-
the
image
is
not
sharp,
the
plate
is
damaged
or
dirty.
Please
correct
all
deficiencies
and
press
Capture
.
After
all
the
necessary
snapshots
are
taken,
wait
for
the
result
of
calibration procedure.
From now on the
resulting accuracy
of
calibration
will be
specified on the home screen
in
pixels
.
A
good
result is accuracy
no worse (not higher) than 0.1 pix
. If you get a higher
result, repeat the calibration, carefully checking the plates position and brightness. If
is
is
necessary, repeat calibration without
blue
light
and
filters.