![background image](http://html1.mh-extra.com/html/rangevision/neo/neo_setup-and-calibration-instructions_3644728014.webp)
The
orientation
can
be
started
from
the
Calibration
window
on the
Start
screen
.
●
place
the
calibration
plate
at
the working
distance
to
the
scanner
(the first
calibration
position),
●
press
the
Capture
button,
●
next,
following
the
steps,
make
two
more
shots.
●
after
this
the
accuracy
of
orientation will
be
shown.
It
should
be approximately
equal
to
the
accuracy value
of
the
last
calibration.
If
the
accuracy
during
orienting
starts
to
be very
different
from
the
original
one,
you
need
to
re-calibrate
the
equipment.
Example
:
The
initial
accuracy
is
0.084,
accuracy
after
orientation
is
0.25.
The
value
exceeded
0.2,
re-calibration
is
required.
Automatic
calibration
The
automatic
calibration
can
be
started
from
the
Calibration
window
on the
Start
screen
,
but
it
is
only
accessible
when
turntable
is
connected
and recognized
by
the
software.
Automatic
calibration
serves
the
same
purpose
as
full
calibration,
but
it
is easier
to
perform,
because
it
uses
turntable
to
shift
the
calibration
plate between
positions.
14
Содержание NEO