background image

Data Sheet-3

AD 22 & AD 22B Block Diagram

Application Information

SENSITIVITY

Adjust the Sensitivity control so its indicator points to the

nominal input signal level. This ensures optimal performance.
Signal present and Clip indicators provide verification that the
signal is within the optimal range. The Sensitivity control
adjusts the input and output signal so that the AD 22 or AD
22B is always at unity gain.

SETTING DELAY

There are two modes for setting Delay in the AD 22 and

AD 22B: setting one channel at a time and setting 

both

channels simultaneously. To set a single channel’s Delay,
press the Channel button until the LEDs indicate the Channel
you want to set (1 or 2). Then press the up/down buttons until
the display shows the desired Delay. When editing 

both

channels simultaneously (both Channel 1 and 2 LEDs 

on

), the

display shows the 

smaller

 of the two current Delay values. In

this mode, the two current Delay values are “locked” together.
Adjusting the up/down buttons changes both values by the
same relative amount. The 

edit both mode

 allows easy stereo

editing and also allows both drivers of a pre-aligned cluster to
be moved forward or backward simultaneously.

STORING DELAY

Press the Store button. This stores 

both

 current Delay

values into each channel's current Memory. A channel's
current Memory is indicated by the Memory LED lit when
editing that channel. The Store LED stops flashing when the
current Delay values match the stored values.

RECALLING DELAY

Pressing the Recall button alternately recalls stored Memories

(A then B then A…) for the selected channel(s) only.

SETTING TEMPERATURE

Hold down Display Mode and press the up/down buttons

to edit the AD 22B’s temperature setting. The 1 ms/Coarse
buttons display temperature in degrees Celsius, the 10µsec/
Fine buttons display degrees in Fahrenheit. No matter which
unit (°C or °F), adjustments are always in 1°C steps (or
1.8°F). The AD 22B does not change the current Delay 

times

for different temperature settings; 

only the displayed 

distance

values are altered.

 

(See example below.)

REMOTE RECALL

A switch wired to the Remote Recall terminals allows

remote recall of stored Memories for each channel. Wire
directly to a room divider latch to automatically recall the two
room configurations. Or store one Memory with the speaker
stack’s distance at one temperature and store the other
Memory with the stack’s distance at another temperature.
Then, during the warmer part of the day, restore the warmer
temperature.

For example, set up the initial temperature for 71.6°F (the

default), and set the stack’s distance for 250.00 feet. Press
Store

 

to save this value in Memory A (250.00 feet at 71.6°F is

220.85 msec). Then change the temperature setting to 100°F,
and we’re in Arizona. Notice that the current distance changes
to 257.11 feet, but the current Delay 

time

 is 

still 220.85 msec

.

Now, edit the current value for 250.00 feet (still your stack’s
distance). Press Store to save this value in Memory B (250.00
feet at 100°F is 214.74 msec.) Now Memory A has the proper
value for the stack at 71.6°F (220.85 msec/250 ft.), and
Memory B has the value for 100°F (214.74 msec/250 ft.).

Отзывы: