Command reference
R&S
®
ZNL/ZNLE
903
User Manual 1178.5966.02 ─ 19
<CalUnitPort1>
Number of the calibration unit port, to whom test port <Test-
Port1> is assigned.
<TestPort2>
Number of the second test port ...
Example:
Prerequisite:
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
'MyCalU'
Sets 'MyCalU' as the active calibration unit.
SYSTem:COMMunicate:RDEVice:AKAL:PORTs? 'abc'
Queries the ports of characterization 'abc'; returns something
like '3,N 50 Ohm,MALE,4,N 50 Ohm,MALE'. In particular, cali-
bration unit ports 1 and 2 are not yet characterized.
SENSe:CORRection:COLLect:AUTO:CKIT:PORTs:ADD
'abc',1,2
Extend characterization 'abc' with test port 1 assigned to port 2
of the cal unit.
SYSTem:COMMunicate:RDEVice:AKAL:PORTs? 'abc'
Now returns something like '2,N 50 Ohm,MALE,3,N 50
Ohm,MALE,4,N 50 Ohm,MALE'.
SENSe:CORRection:COLLect:AUTO:CKIT:PORTs:ADD
'abc',1,2
Repeat characterization of test port 2, e.g. because the connec-
tion cable wasn't fastened correctly ...
Usage:
Setting only
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
<CalType>,
<Characterization>[, <CalName>]
Selects a calibration type and a cal unit characterization (cal kit file) for an automatic
calibration.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Parameters:
<CalType>
FNPort | TOSM | FOPort | OPTPort | FRTRans | FTRans |
RTRans | REFL | RSHort
Calibration Type
FNPort
Full n-port (UOSM) calibration
TOSM
Full n-port calibration with characterized Through.
If the selected cal unit characterization does not contain a
Through characterization, the command silently falls back to a
FNPort (UOSM) calibration.
FOPort
Full one-port ("Refl OSM") calibration
VNA command reference