Command reference
R&S
®
ZNL/ZNLE
719
User Manual 1178.5966.02 ─ 19
Setting parameters:
<SweepType>
FLIN | FLOG | FSEG | FSINgle | TLIN | TLOG
Keywords for the units of the stimulus values; frequency, power
(R&S
ZNL only), and time units.
The selected unit must be compatible with the sweep type (see
FLOG, FSEG and FSINgle, s for TLIN and TLOG, dBm for PLIN,
PLOG and PSINgle. Otherwise the ripple limit lines cannot be
displayed and no ripple limit check is possible.
*RST:
FLIN
Example:
SWE:TYPE LOG
Select a logarithmic frequency sweep.
CALC:RIPP:CONT:DOM FLIN
Delete all existing ripple limit ranges and select linear frequency
units for the domain of the active trace.
CALC:RIPP:DATA 1, 1000000000, 2000000000, 3
Define and enable a ripple limit range in the stimulus range
between 1 GHz and 2 GHz, assigning a ripple limit of 3 dB.
Usage:
Setting only
Manual operation:
See
"Add / Insert / Delete / Delete All / Align All"
CALCulate<Chn>:RIPPle:DATA
<RippleLimRange>...
Adds and enables/disables an arbitrary number of ripple limit ranges, assigning the
stimulus values and the ripple limits. See
Chapter 9.4.1.2, "Rules for ripple test defini-
Note:
This command does not overwrite existing ripple limit ranges. The defined
ranges are appended to the range list as new ranges. Use the
CALCulate<Chn>:RIPPle:SEGMent<Seg>...
commands to change existing ripple
limits.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<RippleLimRange>
Parameter list in the format <Type>, <StartStimulus>, <StopSti-
mulus>, <RippleLimit>[, {<Type>, <StartStimulus>, <StopStimu-
lus>, <RippleLimit>}], where:
<Type>
–
Boolean identifier for the ripple limit range type. 1 for
ripple limit range on (with limit check). 0 for ripple limit range off:
The range is defined, but no limit check result displayed. The
result is still available via
<StartStimulus>/<StopStimulus>
–
stimulus values (unitless) con-
fining the ripple limit range
<RippleLimit>
–
ripple limit (unitless) in the stimulus range
between <StartStimulus> and <StopStimulus>
VNA command reference