VNA GUI reference
R&S
®
ZNL/ZNLE
380
User Manual 1178.5966.02 ─ 19
Remote command:
CALCulate<Chn>:STATistics:NLINear:COMP[:STATe]
CALCulate<Chn>:STATistics:NLINear:COMP:LEVel
CALCulate<Chn>:STATistics:NLINear:COMP:PHASe
CALCulate<Chn>:STATistics:NLINear:COMP:RESult?
CALCulate<Chn>:STATistics[:STATe]:AREA
Reference Value
(R&S
ZNL only)
This control is only available if
statistics are calculated. It selects the
reference value ("small signal value") for the compression point calculation.
"First Point"
(default)
uses the value at the first point of the selected evaluation range as
the reference value (see
Chapter 10.6.8.2, "Evaluation Range dialog"
"Selected
Marker"
uses the value at a selectable marker as the reference value (see
"Range"
uses the average value in a configurable
as the reference value
"Defined
Value"
allows to specify the reference value manually (see
Remote command:
CALCulate<Chn>:STATistics:NLINear:COMP:REFerence
Selected Marker
(R&S
ZNL only)
This button is only visible if "Selected Marker" is used as
. It allows
you to select the marker whose value shall be used as the reference ("small signal
value") for the compression point calculation.
Remote command:
CALCulate<Chn>:STATistics:NLINear:COMP:RMARker
Ref. Range
(R&S
ZNL only)
This button is only visible if "Range" is used as
.
It opens the "Reference Range" dialog that allows you to configure the sweep range
whose average value shall be used as the reference ("small signal value") for the com-
pression point calculation.
Trace softtool