VNA GUI reference
R&S
®
ZNL/ZNLE
311
User Manual 1178.5966.02 ─ 19
●
While in Cartesian diagrams the x-axis represents the stimulus values, in polar and
Smith diagrams this direct relation is lost. In any case, a marker can be used to dis-
play the stimulus value of a given trace point.
For "Time Domain" traces, points in the "observation interval" are interpreted as
stimulus values.
●
All stimulus settings except the "Time Domain" settings are channel settings. "Time
Domain" applies to the active (time domain) trace only.
Background information
Refer to
Chapter 9.1.3, "Traces, channels and diagrams"
10.2.1
Stimulus tab
Defines the sweep range in the current channel, depending on the sweep type.
Related Settings
Refer to the following sections:
●
Chapter 9.1.4.2, "Stimulus and sweep types"
●
Chapter 10.5.2, "Sweep Type tab"
The following "Stimulus" settings are also available on the "Power" tab:
●
●
(R&S
ZNL only)
Start Frequency / Stop Frequency / Center Frequency / Span Frequency
Defines the sweep range for non-segmented frequency sweeps.
sweep, setting "Start Frequency" and "Stop Frequency" or "Center Fre-
quency" and "Span Frequency" are alternatives.
Stimulus softtool