What's new in firmware version 1.70
R&S
®
ZNL/ZNLE
21
User Manual 1178.5966.02 ─ 19
3
What's new in firmware version 1.70
This section lists the changes introduced in the current version of the R&S
ZNL/ZNLE
firmware.
Software version
► To check your R&S
ZNL/ZNLE "Instrument Firmware" and "VNA Server" version,
select [SETUP] > "System Config" > "Ve Options".
Downgrade to a firmware version < 1.22
To downgrade the firmware from a version X
≥
1.22 to a version Y < 1.22, it is required
to uninstall firmware version X manually (using Windows "Apps & Features"). Then,
after a possible reboot of the instrument, install firmware version Y.
For details on manual firmware deinstallation and installation see the annex of the
R&S
ZNL/ZNLE User Manual or Help System.
3.1
Firmware version 1.70
This section lists the changes introduced in firmware version 1.70.
New functionality
●
Power sweep, power meter and power cal support for R&S
ZNL
New remote control functionality
●
New command
MMEMory:STORe:TRACe:OPTion:COMMent
to be added to exported trace files
Improvements
●
For calibration kit standards defined using snp files, it is now possible to export
these files
See
"Write .s1p File/Write .s2p File"
●
Command
DISPlay[:WINDow<Wnd>]:TRACe:EFEed
can now move live and
memory traces between diagram areas
Solved issues
●
With multiple traces and coupled markers in channel
<Chn>
, adding a marker
<no>
greater than 10 using
CALCulate<Chn>:MARKer<no>:STATe ON
did not
work as expected.
●
If measurement data is available for a calibration, and the related setup was saved
under a different name, then "Repeat Calibration" failed in the resulting recall set
●
If measurement data is available for a calibration of channel no. x and this calibra-
tion was stored to the cal pool, then "Repeat Calibration" failed if the related cal
pool calibration was applied to a channel no. y≠x
●
Changing the smoothing aperture did not take effect immediately
Firmware version 1.70