Command Reference
R&S
®
ZNC
630
User Manual 1173.9557.02 ─ 13
Manual operation:
See
"Flatness Cal > Total Readings"
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:AVERage:NTOLerance
<Tolerance>
Specifies the maximum deviation of the measured power from the target power of the
calibration. The command is valid for all channels and calibrated ports.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the toler-
ance value is valid for all sources.
<PhyPt>
Calibrated port number. This parameter is ignored; the tolerance
value is valid for all sources.
Parameters:
<Tolerance>
Tolerance value
Range:
0.001 dB to 1000 dB
*RST:
1 dB
Default unit: dB
Example:
See
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:CFACtor
<Convergence>
Specifies the convergence factor for a source power calibration.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the con-
vergence factor is valid for all sources.
<PhyPt>
Calibrated port number. This parameter is ignored; the conver-
gence factor is valid for all sources.
Parameters:
<Convergence>
Convergence factor
Range:
0 to 2
Increment: 1 dB
*RST:
1
Example:
See
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:PMReadings
<NoReadings>
Selects the number of power meter readings for source power calibration method
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:METHod RRAFter
.
SCPI Command Reference