Command Reference
R&S
®
ZNC
609
User Manual 1173.9557.02 ─ 13
Parameters:
<MeasDelay>
Measurement delay before each partial measurement. Changing
the delay leaves the number of points unchanged but has an
impact on the duration of the sweep (
Range:
0 s to 2500 s
*RST:
0 s
Default unit: s
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default
settings.
SEGM:SWE:DWEL 1 MS
Set the meas. delay in segment no. 1 to 1 ms.
SEGM:DEF? Response:
300000,8000000000,51,-300,0.056559,0,10000,POS,
NORM
Query the channel parameters for sweep segment 1. The
response value for the segment sweep time (olive) implicitly con-
tains the defined meas. delay.
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl:CONTrol
<Boolean>
Selects common or independent "Meas. Delay" settings for the sweep segments.
Suffix:
<Ch>
.
Channel number.
<Seg>
Sweep segment number. This suffix is ignored; the command
controls the whole segmented sweep.
Parameters:
<Boolean>
ON
–
use independent delay settings, to be defined via
OFF
–
reset the delay in all sweep segments to the delay for
unsegmented sweeps, defined via
The parameter is automatically switched to ON when a meas.
delay time is entered using
*RST:
OFF
SCPI Command Reference