Command Reference
R&S
®
ZNC
565
User Manual 1173.9557.02 ─ 13
Example:
*RST; CORR:COLL:FIXT:LMP:LOSS OFF
Configure a fixture compensation measurement (for all channels):
The analyzer performs an Auto Length (no loss) calculation.
CORR:COLL:FIXT:ACQ OPEN, 2; :CORR:COLL:FIXT:ACQ
SHOR, 4
Perform a fixture compensation sweep at port 2, terminated with
an open standard, and at port 4, terminated with a short.
CORR:COLL:FIXT:SAVE
Save and apply the compensation data.
CORR:COLL:FIXT:STAR
Prepare a new fixture compensation measurement, deleting the
previous data for channel 1.
CORR:COLL:FIXT:LMP OFF
Select a Direct Compensation measurement (for all channels and
traces).
CORR:COLL:FIXT:ACQ SHOR, 1, 3
Perform a fixture compensation sweep at ports 1 and 3, terminated
with a short standard.
CALC2:PAR:SDEF 'Trc2', 'S22'
Create channel no. 2 with a trace named Trc2.
SENS2:CORR:COLL:FIXT:STAR
Prepare a fixture compensation measurement for channel 2. The
channel 1 data is not affected.
SENS2:CORR:COLL:FIXT:ACQ SHOR, 1, 3
Repeat the previous fixture compensation sweep for channel 2.
SENS1:CORR:COLL:FIXT:SAVE; :SENS2:CORR:COLL:
FIXT:SAVE
Save and apply the compensation data for both channels.
SENS1:CORR:OFFS3:DFC?; :SENS2:CORR:OFFS3:DFC?
Query whether the analyzer uses Direct Compensation results at
port 3. The response is
1;1
(true for both channels).
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:FIXTure:LMParameter:LOSS[:STATe]
<Boolean>
Selects an Auto Length or an Auto Length and Loss calculation.
Suffix:
<Ch>
.
Channel number of the corrected channel. This suffix is ignored;
the setting is valid for all channels.
Parameters:
<Boolean>
1 - Auto Length and Loss
0 - Auto Length, no loss
*RST:
1
SCPI Command Reference