Command Reference
R&S
®
ZNC
551
User Manual 1173.9557.02 ─ 13
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe
<Boolean>
Activates or deactivates the calibration mode where the raw measurement data of the
standards is stored after the calibration is completed ("Repeat Previous Cal"). The setting
is valid for the current calibration, where it overwrites the global setting (
). A new calibration deletes
the calibration data acquired in previous calibrations.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Parameters:
<Boolean>
ON | OFF - "Repeat Previous Cal" on or off.
*RST:
OFF
Example:
CORR:COLL:RSAV:DEF ON
Generally keep the raw measurement data of the standards after
the calibration is completed.
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard as
calibration type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the raw
measurement results of this standard.
CORR:COLL:RSAV OFF
To save disk space, delete the current raw calibration data after
the calibration is completed.
CORR:COLL:SAVE:SEL
Calculate the system error correction data and apply it to the active
channel.
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe:DEFault
<Boolean>
Activates or deactivates the storage of raw measurement data after the calibration is
completed ("Save Sweep Data"). The setting remains valid for all subsequent calibrations
until it is changed explicitly. A new calibration deletes the calibration data acquired in
previous calibrations.
This setting is automatically disabled and deactivated in case a switch matrix is used in
the RF configuration.
Suffix:
<Ch>
.
Channel number of the calibrated channel.
Parameters:
<Boolean>
ON | OFF - "Save Sweep Data" on or off.
*RST:
OFF
Example:
See
SCPI Command Reference