Command Reference
R&S
®
ZNC
486
User Manual 1173.9557.02 ─ 13
Manual operation:
See
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:X:OFFSet
<StimulusOffset>
Shifts the trace
<WndTr>
in horizontal direction, leaving the positions of all markers
unchanged.
Suffix:
<Wnd>
.
Number of an existing diagram area (defined by means of
ON
).
<WndTr>
Existing trace number, assigned by means of
Parameters:
<StimulusOffset>
Stimulus offset value.
Range:
-1000 GHz to +1000 GHz. The range and unit
depends on the sweep type.
*RST:
0
Default unit: Hz
Example:
*RST; :DISP:WIND:TRAC:X:OFFS 1MHZ; :DISP:WIND:
TRAC:Y:OFFS 10
Create the default trace and shift it horizontally by 1 MHz, vertically
by 10 dB.
Manual operation:
See
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y:OFFSet
<MagnitudeFactor>[,
<PhaseFactor>, <RealPart>, <ImaginaryPart>]
Modifies all points of the trace
<WndTr>
by means of an added and/or a multiplied com-
plex constant. The response values M of the trace are transformed according to:
Suffix:
<Wnd>
.
Number of an existing diagram area (defined by means of
ON
).
<WndTr>
Existing trace number, assigned by means of
Parameters:
<MagnitudeFactor>
Multiplied magnitude factor
Range:
-300 dB to + 300 dB
*RST:
0 dB
Default unit: dB
<PhaseFactor>
Multiplied phase factor, optional for setting command but returned
by query
Range:
-3.4*1038 deg to +3.4*1038 deg
*RST:
0 deg
Default unit: deg
SCPI Command Reference