Command Reference
R&S
®
ZNC
409
User Manual 1173.9557.02 ─ 13
Usage:
Query only
Manual operation:
See
CALCulate<Chn>:MARKer<Mk>:FUNCtion:SPAN
Sets the sweep span of the sweep range equal to the absolute value of the first coordinate
of the active delta marker
<Mk>
on trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10.
Example:
*RST; :CALC:MARK ON; MARK:DELTa ON
Create marker 1 in the center of the current sweep range and
enable the delta mode.
CALC:MARK:X 300MHz
Increase the stimulus value of the delta marker by 300 MHz.
CALC:MARK:FUNC:SPAN
Set the sweep range equal to 300 MHz. The sweep range starts
at the reference marker position, i.e. in the center of the analyzer's
frequency range.
Usage:
Event
Manual operation:
See
"Center / Start / Stop / Span = Marker"
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STARt
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STOP
These command sets the beginning (
...STARt
) and the end (
...STOP
) of the sweep
range equal to the stimulus value of the marker
<Mk>
on trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10
Example:
*RST; :CALC:MARK ON
Create marker 1 in the center of the current sweep range and
assign it to trace no. 1.
CALC:MARK:FUNC:STAR
Divide the sweep range in half, starting at the current marker posi-
tion. As an alternative:
CALC:MARK:FUNC:STOP
Divide the sweep range in half, ending at the current marker posi-
tion.
Usage:
Event
Manual operation:
See
"Center / Start / Stop / Span = Marker"
SCPI Command Reference