GUI Reference
R&S
®
ZNC
126
User Manual 1173.9557.02 ─ 13
Wave Quantity
Selects the type (left pull-down list) and the port number assignment (right pull-down list)
of the wave quantitiy.
Remote command:
"<Trace_Name>", "A1" ...
Create new trace and select trace name and measurement parameter:
"<Trace_Name>", "A1" ...
Source Port
Selects the the source port for the stimulus signal ("Port 1" or "Port 2").
The analyzer places no restriction on the combination of source ports and port numbers
of the measured wave quantity, so it is even possible to measure a
2
while the source port
is Port 1 (e.g. in order to estimate the directivity of the coupler in the internal test set).
Remote command:
"<Trace_Name>", "A1" ...
Create new trace and select trace name and measurement parameter:
"<Trace_Name>", "A1" ...
(the parameter name <Ratio> can also contain the source port)
Show as
Selects the physical unit of the displayed trace. It is possible to display the measured
"Voltage" V or convert the wave quantity into an effective power according to P = V
2
/
Re(Z
0
). Z
0
denotes the reference impedance of the source port (for wave quantities a
n
)
or of the receive port (for wave quantities b
n
).
The reference impedances are defined in the "Balanced Ports" dialog; see
ter 4.2.1.3, "Balanced Ports Dialog"
Remote command:
Detector
Selects the algorithm that is used to calculate the displayed measurement points from
the raw data.
chapter 3.3.4.3, "Detector Settings"
Remote command:
"<Trace_Name>", "<Ratio>"
"<Trace_Name>", "<Ratio>"
(the parameter name <Ratio> can also contain the detector)
4.2.1.8
Meas > Z <-- Sij
Selects converted impedances as measured quantities. The impedances are calculated
from the measured S-parameters.
Trace Settings