Command Reference
R&S
®
ZNC
613
User Manual 1173.9557.02 ─ 13
Parameters:
<Sweep>
Number of consecutive sweeps to be measured.
Range:
1 to 100000
*RST:
1
Example:
See
Manual operation:
See
[SENSe:]SWEep:COUNt:ALL
<Sweep>
Defines the number of sweeps to be measured in single sweep mode (
OFF
). The setting is applied to all channels. Use
to define the sweep count for a single channel.
Setting parameters:
<Sweep>
Number of consecutive sweeps to be measured.
Range:
1 to 100000
*RST:
1
Example:
See
chapter 7.2.4, "Data Handling"
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]SWEep:DETector:TIME
<DetectorTime>
Defines the detector observation time ratios and wave quantities if an "AVG Real Imag"
or an "AVG Mag Phase" detector is active. The detector is selected together with the
measured quantity (
).
Suffix:
<Ch>
.
Channel number
Parameters:
<DetectorTime>
Detector time
Range:
0 s to 3456000 s
*RST:
0.01 s
Default unit: s
Example:
*RST; :CALC:PAR:MEAS 'TRC1', 'a1D1RMS'
Select the wave quantity a
1
for the default trace and activate the
RMS detector.
SWE:DET:TIME 1
Specify an observation time of 1 s at each sweep point.
Manual operation:
See
SCPI Command Reference