GUI Reference
R&S
®
ZNC
119
User Manual 1173.9557.02 ─ 13
●
To dissolves a balanced port, select it and tap "Single".
●
To exclude a single physical port from the measurement, select the port and tap
"Unused".
Remote command:
Reference Impedance
The "Reference Impedance" tab of the "Balanced and Measured Ports" dialog
(re-)defines the port impedances for differential and common mode.
Background information
Refer to
chapter 3.3.5.4, "Reference Impedances"
Access:
E.g. TRACE > MEAS > S-Params > Balanced Ports...
The default reference impedance for a physical port is equal to the reference impedance
of the connector type assigned to the port but can be defined as an arbitrary complex
value (renormalization of port impedances). By changing the reference impedance, it is
possible to convert the measured values at 50 Ω (75 Ω) into values at arbitrary port
impedances.
Trace Settings