Command Reference
R&S
®
ZNB/ZNBT
828
User Manual 1173.9163.02 ─ 55
Parameters:
<Boolean>
1 (ON, true) if the test fixture is connected to port <Ph_pt>, 0
(OFF, false) otherwise
Example:
See
Chapter 9.2.8, "Fixture Modeling"
Manual operation:
See
CALCulate:FMODel:ISD<Ph_pt>:FTIMe:DUT
<FltDutAndLeadIns>
If
CALCulate:FMODel:ISD<Ph_pt>:FTIMe:OVERride
mand allows to set the flight time for DUT + Lead-ins manually.
This is equivalent to setting the
leadin_dut_time
batch mode parameter of the ISD
tool to a numeric value (not
auto
).
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<FltDutAndLeadIns> Default unit: s
Manual operation:
See
CALCulate:FMODel:ISD<Ph_pt>:FTIMe:OVERride
<Boolean>
If set to true, the flight time for DUT + lead-in is specified using
. Otherwise, it is calculated automatically.
True is equivalent to setting the
leadin_dut_time
batch mode parameter of the ISD
tool to
auto
.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<Boolean>
Manual operation:
See
"Automatic Flt Tm for DUT + Lead Ins"
CALCulate:FMODel:ISD<Ph_pt>:OPERation
<Operation>
Defines the execution mode of the ISD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<Operation>
FAST | ACCurate
ACCurate: normal execution mode
FAST: reduces the execution time to ~50%
In many cases the FAST mode can be enabled with only little
loss of accuracy.
SCPI Command Reference