Command Reference
R&S
®
ZNB/ZNBT
1242
User Manual 1173.9163.02 ─ 55
Parameters:
<DetectorTime>
Detector time
Range:
0 s to 3456000 s
*RST:
0.01 s
Default unit: s
Example:
*RST; :CALC:PAR:MEAS 'TRC1', 'a1D1AVG'
Select the wave quantity a
1
for the default trace and activate the
AVG detector.
SWE:DET:TIME 1
Specify an observation time of 1 s at each sweep point.
Manual operation:
See
[SENSe<Ch>:]SWEep:DWELl
<MeasDelay>
Defines the "Meas. Delay" before each partial measurement or the first partial mea-
surement (see
[SENSe<Ch>:]SWEep:DWELl:IPOint
). Setting a delay disables the
automatic calculation of the (minimum) sweep time (see
).
Suffix:
<Ch>
.
Channel number
Parameters:
<MeasDelay>
Measurement delay before each partial measurement. Changing
the delay leaves the number of points unchanged but has an
impact on the duration of the sweep (
Range:
0 s to 13680 s
*RST:
0 s
Default unit: s
Example:
*RST
Reset the instrument, activating a frequency sweep with the S-
parameter S
21
as a measurement result for channel and trace
no. 1.
SWEep:TIME?
Query total sweep time.
SWEep:DWEL 1
SWEep:DWEL:IPOint ALL
Set a delay of 1 s for each partial measurement.
SWE:TIME?
Query total sweep time. The time is extended by the delay times
the total number of sweep points (one partial measurement per
sweep point required).
Manual operation:
See
SCPI Command Reference