Command Reference
R&S
®
ZNB/ZNBT
1180
User Manual 1173.9163.02 ─ 55
This command extends the functionality of manual control, where the reference
receiver calibration is included in the source power calibration of each port (and always
ON).
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the receiver power calibration
for the reference waves a
n
.
*RST:
OFF
Example:
See
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
[SENSe<Ch>:]CORRection:POWer<PhyPt>:IMODulation:ACQuire
[<CalOnlyPortFreq>]
Starts the receiver calibration (3
rd
power calibration step for intermodulation measure-
ments), stores and applies the calibration data.
Note:
The receiver calibration relies on the source power calibration acquired in step
nos. 1 and 2.
[SENSe<Ch>:]CORRection:POWer<PhyPt>:IMODulation:ACQuire
is enabled
only after steps 1 and 2 have been carried out (
CORRection:IMODulation:LTONe[:ACQuire]
CORRection:IMODulation:UTONe[:ACQuire]
).
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Port number. This suffix is ignored; the analyzer calibrates the
port that is selected as a receive port for the intermodulation
measurement
(
[SENSe<Ch>:]FREQuency:IMODulation:RECeiver
).
Setting parameters:
<CalOnlyPortFreq>
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:STIMulus?
[SENSe<Ch>:]CORRection:STIMulus:PORT<PhyPt>?
Queries the stimulus values of the active calibration. A calibration must be selected
before the command is executed; see example.
SCPI Command Reference