Command Reference
R&S
®
ZNB/ZNBT
1152
User Manual 1173.9163.02 ─ 55
Example:
CORR:COLL:AUTO:TYPE FNPort, '', 1, 2, 4
Perform an automatic full 3-port calibration at test ports 1, 2,
and 4 using the calibration unit's default calibration kit file and
arbitrary test ports of the cal unit.
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:CKIT:INSTall
<CalKitFile>, <Gender>[,
<Ports>]
Allows you to load cal kit data by gender:
●
From the given file
●
To the given ports or all ports
Use
on page 1066 to get information about cal kit files.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<CalKitFile>
Path to the cal kit file, either absolute or relative to the current
directory (see
<Gender>
MALE | FEMale | NGENder
<Ports>
Either a comma-separated list of port numbers or ALL (optional)
Usage:
Setting only
[SENSe<Ch>:]CORRection:COLLect:CKIT:LOAD
<CalKitName>, <KitLabel>,
<Gender>[, <Ports>]
Allows you to load cal kit data by name, label and gender:
●
From the pool
●
To the given ports or all ports
Suffix:
<Ch>
.
Channel number
Setting parameters:
<CalKitName>
The name of the cal kit to be loaded
<KitLabel>
The label of the cal kit to be loaded
<Gender>
MALE | FEMale | NGENder
The gender of the cal kit to be loaded
<Port>
Either a comma-separated list of port numbers or ALL (optional)
Usage:
Setting only
SCPI Command Reference